Vol. 98-B, Issue SUPP_10 | Pages 96 - 96
1 May 2016
Uddin F Tayara B Al-Khateeb H Lanting B
Full Access

Background

Fractures of the femoral component are well reported complications that present a challenging task in revision total hip arthroplasty. Albeit being uncommon, with an incidence of 0.23–11%, the consequences can be devastating. Its extraction being a demanding undertaking that is potentially detrimental to the remaining host bone. Several techniques have been described to address this complex issue prior to revision: drilling of the exposed part of the femoral stem and attaching a threaded extraction device, surface undercutting with an extraction device wedged in, femoral trephine techniques, creation of a femoral cortical window, an extended femoral osteotomy procedure, as well as extraction by means of retrograde nail impaction. Here we present the modified technique we employed in the revision of a failed cementless extensively porous coated femoral component that had fractured at the neck-stem interface.

Technique

The proximal femoral component was visualized and an orthopedic burr and a femoral osteotome employed surrounding the component. Utilizing a Midas Rex® MR7 drill with its metal cutting attachment, a circular recess was created in the shoulder of the femoral component. This facilitated the application of the distal end of a universal slap hammer. The component was retrieved successfully with no associated bone loss negating the need for a femoral osteotomy.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 104 - 104
1 May 2016
Petis S Howard J Lanting B Marsh J Vasarhelyi E
Full Access

Introduction

Total hip arthroplasty (THA) is a commonly performed surgical procedure for the treatment of hip arthritis. Approximately 50,000 THAs are performed annually in Canada. The costs incurred to the healthcare system are tremendous, amounting to anywhere between 4.3 and 7.3 billion dollars each year. Despite the substantial financial burden of THA to the Canadian healthcare system, few studies have provided accurate cost estimations of this procedure.

Purpose

To determine the impact of surgical approach on costs of THA from a hospital perspective, and provide an updated cost estimation of THA within a publically funded healthcare system.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 92 - 92
1 Jan 2016
Vandekerckhove P Teeter M Naudie D Howard J MacDonald S Lanting B
Full Access

Summary

The effect of polyethylene wear and lift-off between the tibial and femoral components on the mechanical axis was assessed in primary TKA (Total Knee Arthroplasty) based on retrieval data and full leg radiographs.

Introduction

Controversy exist regarding performing a TKA with component placement in physiologic versus neutral alignment. Some literature indicates good survivorship and superior clinical outcome in undercorrected TKA's for varus osteoarthritic knees. However, other literature indicates decreased survivorship and coronal plane alignment is still one of the contributing factors to wear in total knee arthroplasty (TKA). The two determinants of the intra-articular deformity in TKA arepolyethylene wear and lift-off between the tibial and femoral compartment. The goal of this study was to evaluate the impact of wear and lift-off on the mechanical axis in neutral and varus aligned TKA's.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 94 - 94
1 Jan 2016
Teeter M Lanting B Vasarhelyi E Ivanov T Vandekerckhove P Howard J Naudie D
Full Access

Increased modularity of total hip arthroplasty components has occurred, with theoretical advantages and disadvantages. Recent literature indicates the potential for elevated revision rates of modular neck systems and the potential for metallosis and ALVAL (Aseptic Lymphocyte dominated Vasculitis Associated Lesion) formation at the modular neck/stem site. Retrieval analysis of one modular neck implant design including SEM (Scanning Electron Microscopy) assessment was done and correlated to FEA (Finite Element Analysis) as well as clinical features of patient demographics, implant and laboratory analysis. Correlation of the consistent corrosion locations to FEA indicates that the material and design features of this system may result in a biomechanical reason for failure. The stem aspect of the modular neck/stem junction may be at particular risk.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 96 - 96
1 Jan 2016
Vasarhelyi E Vijayashankar RS Lanting B Howard J Armstrong K Ganapathy S
Full Access

Introduction

Fast track arthroplasty regimens require preservation of motor power to perform early rehabilitation and ensure early discharge (1). Commonly performed nerve blocks like femoral and Sciatic nerve blocks results in motor weakness thereby interfering with early rehabilitation and may also predispose to patient falls (2, 3). Hence, targeting the terminal branches of the femoral and sciatic nerves around the knee joint under ultrasound is an attractive strategy. The nerve supply of interest for knee analgesia are the terminal branches of the femoral nerve, the genicular branches of the lateral cutaneous nerve of thigh, obturator and sciatic nerves (4).

Methods

We modified the performance of the adductor canal block and combined it with US guided posterior pericapsular injection and lateral femoral cutaneous nerve block to provide analgesia around the knee joint. The femoral artery is first traced under the sartorius muscle until the origin of descending geniculate artery and the block is performed proximal to its origin. A needle is inserted in-plane between the Sartorius and rectus femoris above the fascia lata and 5 ml of 0.5% ropivacaine (LA) is injected to block the intermediate cutaneous nerve of thigh. The needle is then redirected to enter the fascia of Sartorius to deliver an additional 5ml of LA to cover the medial cutaneous nerve of thigh following which it is further advanced till the needle tip is seen to lie adjacent to the femoral artery under the Sartorius to perform the adductor canal block with an additional 15–20 ml of LA to cover nerve to vastus medialis, saphenous nerve and posterior division of the obturator nerve (Fig 1). The lateral cutaneous nerve of thigh is optionally blocked with 10 ml of LA near the anterior superior iliac spine between the origin of Sartorius and tensor fascia lata (Fig 2). The terminal branches of sciatic nerve to the knee joint is blocked by depositing 25 ml of local anesthetic solution between the popliteal artery and femur bone at the level of femoral epicondyles (Fig 3).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 93 - 93
1 Jan 2016
Vandekerckhove P Teeter M Naudie D Howard J MacDonald S Lanting B
Full Access

Introduction

Coronal plane alignment is one of the contributing factors to polyethylene wear in total knee arthroplasty (TKA). The goal of this study was to evaluate the wear and damage patterns of retrieved tibial polyethylene inserts in relationship to the overall mechanical alignment and to the position of the tibial component.

Materials and methods

Based on full-length radiographs, ninety-five polyethylene inserts retrieved from primary TKA's with a minimum time in-vivo of five years were analysed for wear and damage. Four alignment groups were compared: valgus, neutral, mild varus and moderate varus. Varus and valgus positioning of the tibial component was analysed for damage score for the neutral and varus aligned groups.