Aims. This study aimed to determine whether lateral femoral wall thickness (LWT) < 20.5 mm was associated with increased revision risk of
Aims. This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures,
Our rural orthopaedic service has undergone service restructure during the COVID-19 pandemic in order to sustain hip fracture care. All adult trauma care has been centralised to the Royal Shrewsbury Hospital for assessment and medical input, before transferring those requiring operative intervention to the Robert Jones and Agnes Hunt Orthopaedic Hospital. We aim to review the impact of COVID-19 on hip fracture workload and service changes upon management of hip fractures. We reviewed our prospectively maintained trust database and National Hip Fracture Database records for the months of March and April between the years 2016 and 2020. Our assessment included fracture pattern (intrascapular vs extracapsular hip fracture), treatment intervention, length of stay and mortality.Aims
Methods
A proximal femur fracture (PFF) is a common orthopaedic presentation, with an incidence of over 25,000 cases reported in the Australian and New Zealand Hip Fracture Registry (ANZHFR) in 2018. Hip fractures are known to have high mortality. The purpose of this study was to determine the utility of the Clinical Frailty Scale (CFS) in predicting 30-day and one-year mortality after a PFF in older patients. A retrospective review of all fragility hip fractures who met the inclusion/exclusion criteria of the ANZHFR between 2017 and 2018 was undertaken at a single large volume tertiary hospital. There were 509 patients included in the study with one-year follow-up obtained in 502 cases. The CFS was applied retrospectively to patients according to their documented pre-morbid function and patients were stratified into five groups according to their frailty score. The groups were compared using Aims
Methods
The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling. A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal Femoral Nail Antirotation (PFNA) implant (Synthes, Oberdorf, Switzerland; n = 6). Group 1 was subjected to unidirectional compression loading (600 N), with an elastomer (70A durometer) replacing loose fracture fragments to simulate surrounding soft-tissue tensioning. Group 2 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading), also with the elastomer replacing loose fracture fragments. Group 3 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading) without the elastomer. All constructs were tested at 2 Hz for 5000 cycles or until cut-out occurred. The medial migration distance (MMD) was recorded at the end of the testing cycles.Objectives
Methods
The Sliding Hip Screw (SHS) is commonly used to treat trochanteric hip fractures. Fixation failure is a devastating complication requiring complex revision surgery. One mode of fixation failure is lag screw cut-out which is greatest in unstable fracture patterns and when the tip-apex distance of the lag screw is > 25 mm. The X-Bolt Dynamic Hip Plating System (X-Bolt Orthopaedics, Dublin, Ireland) is a new device which aims to reduce this risk of cut-out. However, some surgeons have reported difficulty minimising the tip-apex distance with subsequent concerns that this may lead to an increased risk of cut-out. We measured the tip-apex distance from the intra-operative radiographs of 93 unstable trochanteric hip fractures enrolled in a randomised controlled trial (Warwick Hip Trauma Evaluation, WHiTE One trial). Participants were treated with either the sliding hip screw or the X-Bolt dynamic hip plating system. We also recorded the incidence of cut-out in both groups, at a median follow-up time of 17 months.Objectives
Patients and Methods