Acetabular component orientation in total hip arthroplasty (THA)
influences results. Intra-operatively, the natural arthritic acetabulum
is often used as a reference to position the acetabular component.
Detailed information regarding its orientation is therefore essential. The
aim of this study was to identify the acetabular inclination and
anteversion in arthritic hips. Acetabular inclination and anteversion in 65 symptomatic arthritic
hips requiring THA were measured using a computer navigation system.
All patients were Caucasian with primary osteoarthritis (29 men,
36 women). The mean age was 68 years (SD 8). Mean inclination was
50.5° (SD 7.8) in men and 52.1° (SD 6.7) in women. Mean anteversion
was 8.3° (SD 8.7) in men and 14.4° (SD 11.6) in women. Objectives
Methods
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods
We performed A total of 12 cadaveric lower limbs were tested with a commercial
image-free navigation system using trackers secured by bone screws.
We then tested a non-invasive fabric-strap system. The lower limb
was secured at 10° intervals from 0° to 60° of knee flexion and
100 N of force was applied perpendicular to the tibia. Acceptable
coefficient of repeatability (CR) and limits of agreement (LOA)
of 3 mm were set based on diagnostic criteria for anterior cruciate
ligament (ACL) insufficiency.Objectives
Methods