Advertisement for orthosearch.org.uk
Results 1 - 20 of 36
Results per page:
Bone & Joint Open
Vol. 5, Issue 10 | Pages 832 - 836
4 Oct 2024
Kayani B Mancino F Baawa-Ameyaw J Roussot MA Haddad FS

Aims. The outcomes of patients with unexpected positive cultures (UPCs) during revision total hip arthroplasty (THA) and total knee arthroplasty (TKA) remain unknown. The objectives of this study were to establish the prevalence and infection-free implant survival in UPCs during presumed aseptic single-stage revision THA and TKA at mid-term follow-up. Methods. This study included 297 patients undergoing presumed aseptic single-stage revision THA or TKA at a single treatment centre. All patients with at least three UPCs obtained during revision surgery were treated with minimum three months of oral antibiotics following revision surgery. The prevalence of UPCs and causative microorganisms, the recurrence of periprosthetic joint infections (PJIs), and the infection-free implant survival were established at minimum five years’ follow-up (5.1 to 12.3). Results. Of the 297 patients undergoing aseptic revisions, 37 (12.5%) had at least three UPCs obtained during surgery. The UPC cohort included 23 males (62.2%) and 14 females (37.8%), with a mean age of 71.2 years (47 to 82). Comorbidities included smoking (56.8%), hypertension (48.6%), diabetes mellitus (27.0%), and chronic renal impairment (13.5%). The causative microorganisms included Staphylococcus epidermidis (49.6%), Bacillus species (18.9%), Micrococcus species (16.2%), and Cutibacterium acnes (16.2%). None of the study patients with UPCs developed further PJIs or required further surgical intervention during follow-up. Conclusion. The prevalence of UPCs during presumed aseptic revision THA and TKA was 12.5%. The most common causative microorganisms were of low virulence, and included S. epidermidis, Bacillus species, Micrococcus species, and C. acnes. Microorganism-specific antibiotic treatment for minimum three months’ duration of UPCs in presumed aseptic revision arthroplasty was associated with excellent infection-free implant survival at mid-term follow-up. Cite this article: Bone Jt Open 2024;5(10):832–836


Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims. Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up. Methods. This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up. Results. Clinical assessment showed persistent major improvement in all scores. On radiological assessment, only one case showed a lucent line (without symptoms). At last follow-up, wear and migration were below the critical thresholds. No cup-related revisions were needed, indicating an outstanding survival rate of 100%. Conclusion. Isoelastic VEPE cups offer high success rates and may prevent osteolysis, aseptic loosening, and the need for revision surgeries in the long term. However, longer follow-up is needed to validate our findings and confirm the advantages offered by this cup. Cite this article: Bone Jt Open 2024;5(10):825–831


Bone & Joint Open
Vol. 5, Issue 8 | Pages 688 - 696
22 Aug 2024
Hanusrichter Y Gebert C Steinbeck M Dudda M Hardes J Frieler S Jeys LM Wessling M

Aims

Custom-made partial pelvis replacements (PPRs) are increasingly used in the reconstruction of large acetabular defects and have mainly been designed using a triflange approach, requiring extensive soft-tissue dissection. The monoflange design, where primary intramedullary fixation within the ilium combined with a monoflange for rotational stability, was anticipated to overcome this obstacle. The aim of this study was to evaluate the design with regard to functional outcome, complications, and acetabular reconstruction.

Methods

Between 2014 and 2023, 79 patients with a mean follow-up of 33 months (SD 22; 9 to 103) were included. Functional outcome was measured using the Harris Hip Score and EuroQol five-dimension questionnaire (EQ-5D). PPR revisions were defined as an endpoint, and subgroups were analyzed to determine risk factors.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 59 - 65
1 May 2024
Liu WKT Cheung A Fu H Chan PK Chiu KY

Aims

Isolated acetabular liner exchange with a highly crosslinked polyethylene (HXLPE) component is an option to address polyethylene wear and osteolysis following total hip arthroplasty (THA) in the presence of a well-fixed acetabular shell. The liner can be fixed either with the original locking mechanism or by being cemented within the acetabular component. Whether the method used for fixation of the HXLPE liner has any bearing on the long-term outcomes is still unclear.

Methods

Data were retrieved for all patients who underwent isolated acetabular component liner exchange surgery with a HXLPE component in our institute between August 2000 and January 2015. Patients were classified according to the fixation method used (original locking mechanism (n = 36) or cemented (n = 50)). Survival and revision rates were compared. A total of 86 revisions were performed and the mean duration of follow-up was 13 years.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 260 - 268
1 Apr 2024
Broekhuis D Meurs WMH Kaptein BL Karunaratne S Carey Smith RL Sommerville S Boyle R Nelissen RGHH

Aims. Custom triflange acetabular components (CTACs) play an important role in reconstructive orthopaedic surgery, particularly in revision total hip arthroplasty (rTHA) and pelvic tumour resection procedures. Accurate CTAC positioning is essential to successful surgical outcomes. While prior studies have explored CTAC positioning in rTHA, research focusing on tumour cases and implant flange positioning precision remains limited. Additionally, the impact of intraoperative navigation on positioning accuracy warrants further investigation. This study assesses CTAC positioning accuracy in tumour resection and rTHA cases, focusing on the differences between preoperative planning and postoperative implant positions. Methods. A multicentre observational cohort study in Australia between February 2017 and March 2021 included consecutive patients undergoing acetabular reconstruction with CTACs in rTHA (Paprosky 3A/3B defects) or tumour resection (including Enneking P2 peri-acetabular area). Of 103 eligible patients (104 hips), 34 patients (35 hips) were analyzed. Results. CTAC positioning was generally accurate, with minor deviations in cup inclination (mean 2.7°; SD 2.84°), anteversion (mean 3.6°; SD 5.04°), and rotation (mean 2.1°; SD 2.47°). Deviation of the hip centre of rotation (COR) showed a mean vector length of 5.9 mm (SD 7.24). Flange positions showed small deviations, with the ischial flange exhibiting the largest deviation (mean vector length of 7.0 mm; SD 8.65). Overall, 83% of the implants were accurately positioned, with 17% exceeding malpositioning thresholds. CTACs used in tumour resections exhibited higher positioning accuracy than rTHA cases, with significant differences in inclination (1.5° for tumour vs 3.4° for rTHA) and rotation (1.3° for tumour vs 2.4° for rTHA). The use of intraoperative navigation appeared to enhance positioning accuracy, but this did not reach statistical significance. Conclusion. This study demonstrates favourable CTAC positioning accuracy, with potential for improved accuracy through intraoperative navigation. Further research is needed to understand the implications of positioning accuracy on implant performance and long-term survival. Cite this article: Bone Jt Open 2024;5(4):260–268


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims

The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection.

Methods

We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.


Bone & Joint Open
Vol. 4, Issue 11 | Pages 881 - 888
21 Nov 2023
Denyer S Eikani C Sheth M Schmitt D Brown N

Aims

The diagnosis of periprosthetic joint infection (PJI) can be challenging as the symptoms are similar to other conditions, and the markers used for diagnosis have limited sensitivity and specificity. Recent research has suggested using blood cell ratios, such as platelet-to-volume ratio (PVR) and platelet-to-lymphocyte ratio (PLR), to improve diagnostic accuracy. The aim of the study was to further validate the effectiveness of PVR and PLR in diagnosing PJI.

Methods

A retrospective review was conducted to assess the accuracy of different marker combinations for diagnosing chronic PJI. A total of 573 patients were included in the study, of which 124 knees and 122 hips had a diagnosis of chronic PJI. Complete blood count and synovial fluid analysis were collected. Recently published blood cell ratio cut-off points were applied to receiver operating characteristic curves for all markers and combinations. The area under the curve (AUC), sensitivity, specificity, and positive and negative predictive values were calculated.


Bone & Joint Open
Vol. 4, Issue 8 | Pages 551 - 558
1 Aug 2023
Thomas J Shichman I Ohanisian L Stoops TK Lawrence KW Ashkenazi I Watson DT Schwarzkopf R

Aims

United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular tapered stems. No study has evaluated the use of monoblock fluted tapered titanium stems for this indication. This study aimed to evaluate the effects of a monoblock stems on implant survivorship, postoperative outcomes, radiological outcomes, and osseointegration following treatment of THA UCS B2 and B3 periprosthetic fractures.

Methods

A retrospective review was conducted of all patients who underwent revision THA (rTHA) for periprosthetic UCS B2 and B3 periprosthetic fracture who received a single design monoblock fluted tapered titanium stem at two large, tertiary care, academic hospitals. A total of 72 patients met inclusion and exclusion criteria (68 UCS B2, and four UCS B3 fractures). Primary outcomes of interest were radiological stem subsidence (> 5 mm), radiological osseointegration, and fracture union. Sub-analysis was also done for 46 patients with minimum one-year follow-up.


Bone & Joint Open
Vol. 4, Issue 8 | Pages 559 - 566
1 Aug 2023
Hillier DI Petrie MJ Harrison TP Salih S Gordon A Buckley SC Kerry RM Hamer A

Aims. The burden of revision total hip arthroplasty (rTHA) continues to grow. The surgery is complex and associated with significant costs. Regional rTHA networks have been proposed to improve outcomes and to reduce re-revisions, and therefore costs. The aim of this study was to accurately quantify the cost and reimbursement for a rTHA service, and to assess the financial impact of case complexity at a tertiary referral centre within the NHS. Methods. A retrospective analysis of all revision hip procedures was performed at this centre over two consecutive financial years (2018 to 2020). Cases were classified according to the Revision Hip Complexity Classification (RHCC) and whether they were infected or non-infected. Patients with an American Society of Anesthesiologists (ASA) grade ≥ III or BMI ≥ 40 kg/m. 2. are considered “high risk” by the RHCC. Costs were calculated using the Patient Level Information and Costing System (PLICS), and remuneration based on Healthcare Resource Groups (HRG) data. The primary outcome was the financial difference between tariff and cost per patient episode. Results. In all, 199 revision episodes were identified in 168 patients: 25 (13%) least complex revisions (H1); 110 (55%) complex revisions (H2); and 64 (32%) most complex revisions (H3). Of the 199, 76 cases (38%) were due to infection, and 78 patients (39%) were “high risk”. Median length of stay increased significantly with case complexity from four days to six to eight days (p = 0.006) and for revisions performed for infection (9 days vs 5 days; p < 0.001). Cost per episode increased significantly between complexity groups (p < 0.001) and for infected revisions (p < 0.001). All groups demonstrated a mean deficit but this significantly increased with revision complexity (£97, £1,050, and £2,887 per case; p = 0.006) and for infected failure (£2,629 vs £635; p = 0.032). The total deficit to the NHS Trust over two years was £512,202. Conclusion. Current NHS reimbursement for rTHA is inadequate and should be more closely aligned to complexity. An increase in the most complex rTHAs at major revision centres will likely place a greater financial burden on these units. Cite this article: Bone Jt Open 2023;4(8):559–566


Bone & Joint Open
Vol. 4, Issue 7 | Pages 472 - 477
1 Jul 2023
Xiang W Tarity TD Gkiatas I Lee H Boettner F Rodriguez JA Wright TM Sculco PK

Aims. When performing revision total hip arthroplasty using diaphyseal-engaging titanium tapered stems (TTS), the recommended 3 to 4 cm of stem-cortical diaphyseal contact may not be available. In challenging cases such as these with only 2 cm of contact, can sufficient axial stability be achieved and what is the benefit of a prophylactic cable? This study sought to determine, first, whether a prophylactic cable allows for sufficient axial stability when the contact length is 2 cm, and second, if differing TTS taper angles (2° vs 3.5°) impact these results. Methods. A biomechanical matched-pair cadaveric study was designed using six matched pairs of human fresh cadaveric femora prepared so that 2 cm of diaphyseal bone engaged with 2° (right femora) or 3.5° (left femora) TTS. Before impaction, three matched pairs received a single 100 lb-tensioned prophylactic beaded cable; the remaining three matched pairs received no cable adjuncts. Specimens underwent stepwise axial loading to 2600 N or until failure, defined as stem subsidence > 5 mm. Results. All specimens without cable adjuncts (6/6 femora) failed during axial testing, while all specimens with a prophylactic cable (6/6) successfully resisted axial load, regardless of taper angle. In total, four of the failed specimens experienced proximal longitudinal fractures, three of which occurred with the higher 3.5° TTS. One fracture occurred in a 3.5° TTS with a prophylactic cable yet passed axial testing, subsiding < 5 mm. Among specimens with a prophylactic cable, the 3.5° TTS resulted in lower mean subsidence (0.5 mm (SD 0.8)) compared with the 2° TTS (2.4 mm (SD 1.8)). Conclusion. A single prophylactic beaded cable dramatically improved initial axial stability when stem-cortex contact length was 2 cm. All implants failed secondary to fracture or subsidence > 5 mm when a prophylactic cable was not used. A higher taper angle appears to decrease the magnitude of subsidence but increased the fracture risk. The fracture risk was mitigated by the use of a prophylactic cable. Cite this article: Bone Jt Open 2023;4(7):472–477


Bone & Joint Open
Vol. 4, Issue 4 | Pages 226 - 233
1 Apr 2023
Moore AJ Wylde V Whitehouse MR Beswick AD Walsh NE Jameson C Blom AW

Aims

Periprosthetic hip-joint infection is a multifaceted and highly detrimental outcome for patients and clinicians. The incidence of prosthetic joint infection reported within two years of primary hip arthroplasty ranges from 0.8% to 2.1%. Costs of treatment are over five-times greater in people with periprosthetic hip joint infection than in those with no infection. Currently, there are no national evidence-based guidelines for treatment and management of this condition to guide clinical practice or to inform clinical study design. The aim of this study is to develop guidelines based on evidence from the six-year INFection and ORthopaedic Management (INFORM) research programme.

Methods

We used a consensus process consisting of an evidence review to generate items for the guidelines and online consensus questionnaire and virtual face-to-face consensus meeting to draft the guidelines.


Bone & Joint Open
Vol. 3, Issue 12 | Pages 991 - 997
23 Dec 2022
McPherson EJ Stavrakis AI Chowdhry M Curtin NL Dipane MV Crawford BM

Aims. Large acetabular bone defects encountered in revision total hip arthroplasty (THA) are challenging to restore. Metal constructs for structural support are combined with bone graft materials for restoration. Autograft is restricted due to limited volume, and allogenic grafts have downsides including cost, availability, and operative processing. Bone graft substitutes (BGS) are an attractive alternative if they can demonstrate positive remodelling. One potential product is a biphasic injectable mixture (Cerament) that combines a fast-resorbing material (calcium sulphate) with the highly osteoconductive material hydroxyapatite. This study reviews the application of this biomaterial in large acetabular defects. Methods. We performed a retrospective review at a single institution of patients undergoing revision THA by a single surgeon. We identified 49 consecutive patients with large acetabular defects where the biphasic BGS was applied, with no other products added to the BGS. After placement of metallic acetabular implants, the BGS was injected into the remaining bone defects surrounding the new implants. Patients were followed and monitored for functional outcome scores, implant fixation, radiological graft site remodelling, and revision failures. Results. Mean follow-up was 39.5 months (36 to 71), with a significant improvement in post-revision function compared to preoperative function. Graft site remodelling was rated radiologically as moderate in 31 hips (63%) and strong in 12 hips (24%). There were no cases of complete graft site dissolution. No acetabular loosening was identified. None of the patients developed clinically significant heterotopic ossification. There were twelve reoperations: six patients developed post-revision infections, three experienced dislocations, two sustained periprosthetic femur fractures, and one subject had femoral component aseptic loosening. Conclusion. Our series reports bone defect restoration with the sole use of a biphasic injectable BGS in the periacetabular region. We did not observe significant graft dissolution. We emphasize that successful graft site remodelling requires meticulous recipient site preparation. Cite this article: Bone Jt Open 2022;3(12):991–997


Bone & Joint Open
Vol. 3, Issue 11 | Pages 859 - 866
4 Nov 2022
Diesel CV Guimarães MR Menegotto SM Pereira AH Pereira AA Bertolucci LH Freitas EC Galia CR

Aims. Our objective was describing an algorithm to identify and prevent vascular injury in patients with intrapelvic components. Methods. Patients were defined as at risk to vascular injuries when components or cement migrated 5 mm or more beyond the ilioischial line in any of the pelvic incidences (anteroposterior and Judet view). In those patients, a serial investigation was initiated by a CT angiography, followed by a vascular surgeon evaluation. The investigation proceeded if necessary. The main goal was to assure a safe tissue plane between the hardware and the vessels. Results. In ten at-risk patients undergoing revision hip arthroplasty and submitted to our algorithm, six were recognized as being high risk to vascular injury during surgery. In those six high-risk patients, a preventive preoperative stent was implanted before the orthopaedic procedure. Four patients needed a second reinforcing stent to protect and to maintain the vessel anatomy deformed by the intrapelvic implants. Conclusion. The evaluation algorithm was useful to avoid blood vessels injury during revision total hip arthroplasty in high-risk patients. Cite this article: Bone Jt Open 2022;3(11):859–866


Bone & Joint Research
Vol. 11, Issue 9 | Pages 608 - 618
7 Sep 2022
Sigmund IK Luger M Windhager R McNally MA

Aims

This study evaluated the definitions developed by the European Bone and Joint Infection Society (EBJIS) 2021, the International Consensus Meeting (ICM) 2018, and the Infectious Diseases Society of America (IDSA) 2013, for the diagnosis of periprosthetic joint infection (PJI).

Methods

In this single-centre, retrospective analysis of prospectively collected data, patients with an indicated revision surgery after a total hip or knee arthroplasty were included between 2015 and 2020. A standardized diagnostic workup was performed, identifying the components of the EBJIS, ICM, and IDSA criteria in each patient.


Bone & Joint Open
Vol. 3, Issue 6 | Pages 485 - 494
13 Jun 2022
Jaubert M Le Baron M Jacquet C Couvreur A Fabre-Aubrespy M Flecher X Ollivier M Argenson J

Aims. Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. Methods. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test. Results. Our staged exchange practice shows poor results, which means a 42% mechanical spacer failure rate, and a 20% recurrent infection rate over the two years average follow-up period. Moreover, we found a high rate of spacer dislocation (23%) and a low rate of spacer fracture (8%) compared to the previous literature. Our findings stress that the majority of spacer complications and failures is reflecting a population with high comorbid burden, highlighted by the American Society of Anesthesiology grade, Charlson Comorbidity Index, and Lee score associations, as well as the cardiac, pulmonary, kidney, or hepatic chronic conditions. Conclusion. Our experience of a two-stage hip exchange revision noted important complication rates associated with high failure rates of polymethylmethacrylate spacers. These findings must be interpreted in the light of the patient’s comorbidity profiles, as the elective population for staged exchange has an increasing comorbid burden leading to poor results. In order to provide better results for this specific population, our conclusion suggests that comparative strategy studies are required to improve our therapeutic indication. Cite this article: Bone Jt Open 2022;3(6):485–494


Bone & Joint Open
Vol. 3, Issue 5 | Pages 423 - 431
1 May 2022
Leong JWY Singhal R Whitehouse MR Howell JR Hamer A Khanduja V Board TN

Aims

The aim of this modified Delphi process was to create a structured Revision Hip Complexity Classification (RHCC) which can be used as a tool to help direct multidisciplinary team (MDT) discussions of complex cases in local or regional revision networks.

Methods

The RHCC was developed with the help of a steering group and an invitation through the British Hip Society (BHS) to members to apply, forming an expert panel of 35. We ran a mixed-method modified Delphi process (three rounds of questionnaires and one virtual meeting). Round 1 consisted of identifying the factors that govern the decision-making and complexities, with weighting given to factors considered most important by experts. Participants were asked to identify classification systems where relevant. Rounds 2 and 3 focused on grouping each factor into H1, H2, or H3, creating a hierarchy of complexity. This was followed by a virtual meeting in an attempt to achieve consensus on the factors which had not achieved consensus in preceding rounds.


Bone & Joint Open
Vol. 3, Issue 3 | Pages 229 - 235
11 Mar 2022
Syam K Unnikrishnan PN Lokikere NK Wilson-Theaker W Gambhir A Shah N Porter M

Aims

With increasing burden of revision hip arthroplasty (THA), one of the major challenges is the management of proximal femoral bone loss associated with previous multiple surgeries. Proximal femoral arthroplasty (PFA) has already been popularized for tumour surgeries. Our aim was to describe the outcome of using PFA in these demanding non-neoplastic cases.

Methods

A retrospective review of 25 patients who underwent PFA for non-neoplastic indications between January 2009 and December 2015 was undertaken. Their clinical and radiological outcome, complication rates, and survival were recorded. All patients had the Stanmore Implant – Modular Endo-prosthetic Tumour System (METS).


Bone & Joint Research
Vol. 10, Issue 12 | Pages 780 - 789
1 Dec 2021
Eslam Pour A Lazennec JY Patel KP Anjaria MP Beaulé PE Schwarzkopf R

Aims

In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement.

Methods

Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data.


Bone & Joint Open
Vol. 2, Issue 10 | Pages 858 - 864
18 Oct 2021
Guntin J Plummer D Della Valle C DeBenedetti A Nam D

Aims. Prior studies have identified that malseating of a modular dual mobility liner can occur, with previous reported incidences between 5.8% and 16.4%. The aim of this study was to determine the incidence of malseating in dual mobility implants at our institution, assess for risk factors for liner malseating, and investigate whether liner malseating has any impact on clinical outcomes after surgery. Methods. We retrospectively reviewed the radiographs of 239 primary and revision total hip arthroplasties with a modular dual mobility liner. Two independent reviewers assessed radiographs for each patient twice for evidence of malseating, with a third observer acting as a tiebreaker. Univariate analysis was conducted to determine risk factors for malseating with Youden’s index used to identify cut-off points. Cohen’s kappa test was used to measure interobserver and intraobserver reliability. Results. In all, 12 liners (5.0%), including eight Stryker (6.8%) and four Zimmer Biomet (3.3%), had radiological evidence of malseating. Interobserver reliability was found to be 0.453 (95% confidence interval (CI) 0.26 to 0.64), suggesting weak inter-rater agreement, with strong agreement being greater than 0.8. We found component size of 50 mm or less to be associated with liner malseating on univariate analysis (p = 0.031). Patients with malseated liners appeared to have no associated clinical consequences, and none required revision surgery at a mean of 14 months (1.4 to 99.2) postoperatively. Conclusion. The incidence of liner malseating was 5.0%, which is similar to other reports. Component size of 50 mm or smaller was identified as a risk factor for malseating. Surgeons should be aware that malseating can occur and implant design changes or changes in instrumentation should be considered to lower the risk of malseating. Although further follow-up is needed, it remains to be seen if malseating is associated with any clinical consequences. Cite this article: Bone Jt Open 2021;2(10):858–864


Bone & Joint Research
Vol. 10, Issue 9 | Pages 611 - 618
27 Sep 2021
Ali E Birch M Hopper N Rushton N McCaskie AW Brooks RA

Aims

Accumulated evidence indicates that local cell origins may ingrain differences in the phenotypic activity of human osteoblasts. We hypothesized that these differences may also exist in osteoblasts harvested from the same bone type at periarticular sites, including those adjacent to the fixation sites for total joint implant components.

Methods

Human osteoblasts were obtained from the acetabulum and femoral neck of seven patients undergoing total hip arthroplasty (THA) and from the femoral and tibial cuts of six patients undergoing total knee arthroplasty (TKA). Osteoblasts were extracted from the usually discarded bone via enzyme digestion, characterized by flow cytometry, and cultured to passage three before measurement of metabolic activity, collagen production, alkaline phosphatase (ALP) expression, and mineralization.