Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

SLIDING EXTERNAL FIXATOR “A PRODUCT DESIGN”

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress, 2015. PART 3.



Abstract

Background

External fixation for a fracture-dislocation to a joint like the elbow, while maintaining joint mobility is currently done after identifying the center of rotation under X-ray guidance, when applying either a mono-lateral or a circular fixator.

Current treatment

using the galaxy fixation system by Orthofix, the surgeon needs to correctly identify the center of rotation of the elbow under X-ray guidance on lateral views. If the center of rotation of the fixator is not aligned with that of the elbow joint, the assembly will not work, i.e. the elbow will be disrupted on trying to achieve flexion or extension movements.

Figures (A, B, C and D) summarize the critical steps in identifying the centre of rotation (Courtesy of Orthofix Orthopedics International).

New design

This new idea aims to propagate the principle of sliding external fixation applied on the extensor side of a joint, with the limbs of the fixator being able to slide in and out during joint extension and flexion respectively, without hindering the joint movement. Taking the ulno-humeral joint as an example, it is enough to apply the sliding external fixator in line with the subcutaneous border of the ulna, and the pins in the sagital plane, without the need to use x-ray guidance to identify the center of rotation, which simplifies the procedure, and makes it less technically demanding.

The sliding external fixator over the elbow involves two bars which accommodate half pins fixation with headless grip screws to hold the pins, identical to the Rancho cubes technique by Smith & Nephew, these slide snugly into sleeves, those sleeves linked together through a hinge behind the elbow, and the bars are spring loaded to the hinge through the inside of the sleeves, which means they will slid into the sleeves in extension and out in flexion. Length of the sleeve should prevent the bars from dislodgement, and the cross section of both the bars and the sleeves have to correspond to each other for the sleeves to accommodate the bars within them and to prevent rotational instability within the construct itself.

Summary

Applying an external fixator on the extensor surface is an idea could lead to major changes in external fixation product design, the ulno-humeral joint is taken as an example, and other joints could also be addressed taking in consideration joint size and anatomical structures at risk. The sliding technique makes the application easier, without the need to X-ray guided identification of the center of rotation.


*Email: