Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

THE IMPACT OF TIBIAL COMPONENT OVERHANG ON OUTCOME SCORES AND PAIN IN TOTAL KNEE REPLACEMENT

Glasgow Meeting of Orthopaedic Research (GLAMOR)



Abstract

When performing total knee replacement (TKR), surgeons must select a size of tibial component tray that most closely matches the anatomy of the proximal tibia. As implants are available in a limited range of sizes, it may be necessary to slightly under or oversize the component. There are concerns overhang could lead to pain from irritation of soft tissues, and underhang could lead to subsidence and failure.

154 TKRs at 1- or 5-year follow up were reviewed prospectively. Oxford Knee Score (OKS), WOMAC and SF-12 was recorded along with pain scores. Scaled radiographs were reviewed and grouped into perfect sizing (78 TKRs, 50.6%), underhang in isolation (48 TKRs, 31.1%), minor overhang 1–3 mm (10 TKRs, 6.49%) or major overhang >3 mm (18 TKRs, 11.7%).

There was no significant difference in the SF-12 (p=0.356), post-operative OKS (p=0.401) or WOMAC (p=0.466) score. For the OKS, there was no difference for the scores collected at 1 year (p=0.176) or at 5 years (p=0.883).

Pre-operative OKS was well matched between the groups (p=0.152). There was no significant difference in the improvement in OKS from pre-operative scores (p=0.662). There was no significant difference in either the OKS or WOMAC pain scores (p=0.237 and 0.542 respectively).

There was no significant association of medial overhang with?medial knee pain (p=1.000) or lateral overhang with lateral knee pain (p=0.569) when compared to the group of patients with a well sized tibial component.

Our results suggest that tibial component overhang or underhang has no detrimental affect on outcome or pain scores. Surgeons should continue to select the tibial component that most closely fits the rim of the proximal tibia while accepting slight overhang if necessary due to the potential longer-term complications of subsidence and premature failure with an undersized tibial tray.