Abstract
Aim:
To evaluate the effect of a stiffer rod in normalising thoracic hypokyphosis in adolescent idiopathic scoliosis (AIS).
Methods:
A retrospective review of AIS cases performed at our institution was carried out. In order to reduce variability, the analysis included only Lenke 1 cases which had all pedicle screw constructs, with similar constructs and implant density. Cases that underwent anterior release were excluded. All cases had the same implant (Expedium 5.5, Depuy-Synthes, Raynham, USA). The rod material differed in that some cases had 5.5 titanium, whilst others had 5.5 cobalt chrome. The preoperative and postoperative sagittal Cobb angle was measured.
Results:
35 patients met the inclusion criteria. 15 had titanium rods and 20 had cobalt chrome rods. The mean fulcrum correction index was similar between groups. The preoperative coronal and sagittal Cobb was similar between the two groups. There was no statistically significant difference in the postoperative sagittal Cobb between the two groups (ANOVA one way test).
Discussion:
Despite the theoretical advantage of a stiffer construct improving the sagittal profile in AIS, this study identified no such benefit despite closely matching the two groups. All pedicle screw constructs do not seem to improve the sagittal profile despite the use of a stiffer rod.
Conflict of interest:
Depuy-Synthes fund a fellow in this unit.
Conflict Of Interest Statement: No conflict of interest.