Abstract

Introduction:

Restoration of normal hip biomechanics is vital for success of total hip arthroplasty (THA). This requires accurate placement of implants and restoration of limb length and offset. The purpose of this study was to assess the accuracy of computer navigation system in predicting cup placement and restoring limb length and offset.

Material and Methods:

An analysis of 259 consecutive patients who had THA performed with imageless computer navigation system was carried out. Acetabular cup abduction and anteversion, medialisation or lateralisation of offset and limb length change were compared between navigation measurements and follow-up radiographs. Sensitivity, specificity, accuracy and PPV were calculated to assess navigation for cup orientation and student t-test used for evaluation of offset and limb length change.

Results:

Mean cup abduction and anteversion was 40.35° (SD, 5.81) and 18.46° (SD, 6.79) in postop radiographs compared to 41° (SD, 5.03) and 14.76°(SD, 6.11) for navigation measurements. Intraoperative navigation measurements had high PPV and specificity for determining cup abduction and anteversion (PPV >95%, specificity >90%). Accuracy for determining cup abduction was 96.13% compared to 72.2% for cup anteversion. Change in limb length and offset was mean 6.46 mm (SD, 5.68) and −1.07 mm (SD, 5.75) on radiograph evaluation and 5.41 mm (SD, 5.11) and −2.24 mm (SD, 5.87) from navigation measurements respectively, the difference being not significant in both (p value >0.2). Radiograph and navigation had a mean difference of 1.01 mm (SD, 2.83) for offset measurements and a mean difference of 1.05 mm (SD, 4.37) for postop limb length assessment.