Abstract
Hip simulator studies have shown reduced hip offset can cause microseparation and increased wear in hard-on-hard hip bearings. However this has not been analysed yet in vivo. We studied the effect of reduced hip offset on serum metal ion levels in patients with metal-on-metal (MoM) hip arthroplasty.
From all patients who underwent unilateral MoM bearing hip arthroplasty between 2005 and 2009, 63 patients had complete clinical evaluation, measurement of serum chromium and cobalt ion levels as well as biomechanical measurements on pre- and post operative radiographs (cup inclination, head inclination, change in hip offset and change in hip length.)
Ten arthroplasties were revised due to adverse metal reaction and six patients awaiting revision. 55% of ASR hips showed higher metal ions (>7 ppb) whereas only 15% of non-ASR hips had higher ion levels. Patients with reduced postoperative hip offset by more than 5 mm had significantly higher mean metal ion levels compared to the the rest of the hips (31.8 ppb vs. 7.4 ppb, p=0.002). On subgroup analysis this effect was present in non-ASR hips (18.7 ppb vs. 4.7 ppb, p=0.025) but was not significant in ASR hips (29.6 ppb vs. 16.3 ppb, p=0.347).
Our study demonstrated significantly higher serum metal ion levels in patients who lost more than 5 mm hip offset after arthroplasty. Reduced soft tissue tension leading to microseparation of the articulation and edge loading is a theoretical explanation for this effect. This may be relevant in other hard bearings such as ceramic-on-ceramic as well.