ASR Mid-Term Results in a Single Centre: Clinical, Haematological and Radiological Results at More Than Six Years

International Society for Technology in Arthroplasty (ISTA)



Abstract

Introduction:

We present the mid-term results of our consecutive series of 155 hips treated with ASR XL Acetabular System (ASR-XL) for large-diameter metal-on-metal total hip arthroplasty and with DePuy ASR Hip Resurfacing System (ASR) for hip resurfacing.

Methods:

114 ASR-XL and 41 ASR were implanted between 2004 and 2008 in 145 patients (69 men and 76 women) with a mean age of 57 years. Twenty-one patients (23 hips) resulted lost to follow-up. All patients were recalled and monitored periodically with clinical, hematological and radiological evaluation. Average follow-up of the 155 hips was 72 months (1 to 104). The mean follow-up, excluding revised patients and those lost to follow-up, was up to 89 months.

Results:

At latest follow-up we had revised 54 of 155 hips (35%) with complete removal of ASR device. Revision involved 15 hips with ASR (36.5% within ASR group) and 39 hips with ASR-XL (34.2% within ASR-XL group). Main reasons for revision were aseptic loosening in 24 hips (44%), pain or pain associated with elevation of blood metal ions in 9 (17%), elevation of blood metal ions without pain in 9 (17%), deep infection in 4 hips (7%), recurrent dislocation in 1 hip (2%), periprosthetic fractures in 1 hip (2%), unknown because revised elsewhere in 6 (11%). The cumulative survival with revision for any reason as the end-point for ASR and ASR XL were respectively 59.6% and 59.3%.

For patients who did not undergone revision, the mean Harris hip score improved to 91 (57 to 100) at five years and the mean satisfaction after the operation was graded 4.4 in a score from 1 to 5. Metal ions plasma concentration analysis was conducted in 83 patients (87 hips). Elevated metal ion levels (>7 μg/l) was found in 39 patients (42 hips, 48%) with average Cr and Co concentrations respectively of 37.3 μg/l and 81.5 μg/l. The remaining 44 patients (45 hips, 52%) had an average Cr and Co concentrations respectively of 1.2 μg/l and 1.9 μg/l. Metal ion levels in revised patients were significantly (p < 0.001) higher than in non-revised patients.

MARS-MRI performed in 38 hips with high metal ions revealed pseudotumour formations in 13 hips (34%). No macroscopic evident clinical sign of cobaltism was reported for any patient.

Conclusion:

ASR implants survival could have a considerable drop in the mid-term follow-up due to adverse reactions to high blood metal ions concentrations and metal debris. Adverse reactions could include periprosthetic pseudotumour formations and low-detectable periprosthetic osteolysis even in asymptomatic patients.