Abstract
Introduction
The position and orientation of the lower extremities are fundamental for planning and follow-up imaging after arthroplasty and lower extremity osteotomy. But no studies have reported the reproducibility of measurements over time in the same patient, and experience shows variability of the results depending on the protocols for patient positioning. This study explores the reproducibility of measurements in the lower extremity with the patients in “comfortable standing position” by the EOS® imaging system.
Materials and Methods
Two whole-body acquisitions were performed in each of 40 patients who were evaluated for a spine pathology. The average interval between acquisitions was 15 months (4–35 months). Patients did not have severe spine pathology and did not undergo any surgery between acquisitions. The “comfortable standing position” is achieved without imposing on the patient any specific position of the lower limbs and pelvis. All the measurements were performed and compared in both 2- and 3-dimensional images. Distances between the centers of the femoral heads and between the centers of the knees and ankles were measured from the front. The profile is shown by the flexion angle between the axis of the femur (center of the femoral head and the top of the line Blumensaat) and the axis of the tibia.
Results
The average radiation dose was 0.80 mGy (0.5–1.11). For the first acquisition, the mean distance between the femoral heads was 17.9 cm (15.8–20.2), the mean distance between the middle of the knee joints was 16.7 cm (11.2–23.1) and the mean distance between the medial malleoli was 13.1 cm (0 to 18). For the second acquisition, the mean distance between the femoral heads was 17.9 cm (14.9–21.5), the mean distance between the middle of the knee joints was 16.9 cm (11.4–23.1) and the mean distance between the medial malleoli was 13.6 cm (0–19.4). For all comparisons no significant difference was demonstrated in related samples by Wilcoxon rank test and paired Student t test.
Discussion
Two- and 3-dimensional data are not affected by repeated acquisitions several to many months apart in “comfortable standing position.” This work shows the reproducibility of measurements of the lower extremity in the “comfortable standing position” by the EOS® imaging system. Additional research should be considered for combined measures in the face-profile position of each patient.