Abstract
PURPOSE:
Wright Medical has a long history of modular neck hip implants but had fracture issues with the original titanium necks. They subsequently changed to chrome cobalt modular necks. Direct contact between these dissimilar metal parts in the modular femoral component brings into consideration the possibility of similar adverse reactions of metal-on-metal articulations that have been previously described in other designs.
METHODS:
A retrospective review of 10 patients with Wright Medical chrome cobalt modular necks who were evaluated with chromium and cobalt metal ion levels as well as Metal Artifact Reduction Sequence (MARS) MRI's was performed. Pseudotumors were classified by MRI based on wall thickness, T1/T2 signal, shape, and location and given a corresponding type of I, II, or III. For each patient, symptoms or lack thereof were recorded, and time since surgery noted.
RESULTS:
Of 10 patients tested, 9 were symptomatic, and 1 was asymptomatic. The patient that was asymptomatic at last clinical visit at 14 months post-op while symptomatic patients averaged 18 months since initial surgery before symptoms began. Those with metal-poly articulation had an average cobalt level of 1.6, ceramic-ceramic articulation had level of <1, and metal-on-metal had level of 2.9. Five patients had pseudotumor by MRI (2 type I, 1 type II, and 2 type III pseudotumors).
CONCLUSION:
It appears that an unintended consequence of changing from titanium to chrome cobalt modular neck may be occurring secondary to corrosion at neck-stem junction.
SIGNIFICANCE: This reaction does not appear to be design-specific as these findings are similar to our findings in Stryker Rejuvenate stems. Surgeons evaluating patients with these and other similar stems should be aware of this complication and consider ion testing and MARS MRI.