Abstract
Introduction:
In an attempt to reduce stress shielding in the proximal femur multiple new shorter stem design have become available. We investigated the load to fracture of a new polished tapered cemented short stem in comparison to the conventional polished tapered Exeter stem.
Method:
A total of forty-two stems, twenty-one short stems and twenty-one conventional stems both with three different offsets were cemented in a composite sawbone model and loaded to fracture.
Results:
study showed that femurs will break at a significantly lower load to failure with a shorter compared to conventional length Exeter stem.
Conclusion:
This Both standard and short stem design are safe to use as the torque to failure is 7–10 times as much as the torques seen in activities of daily living.