Abstract
Purpose:
To compare accuracy of transepicondylar axis as a reference for femoral component rotation in primary navigated versus non navigated total knee arthroplasty in severely deformed knees.
Methods:
A prospective study done from dec 2009 to dec 2011 at tertiary centre. 180 knees were included (124 females and 56 males). All cases were randomly allocated into 2 groups: navigated and non navigated. All surgeries were carried out by two senior arthroplasty surgeons. All patients undergoing primary total knee replacement were included and all revision cases were excluded. Intraoperative assessment of TEA was done by palpating most prominent point on lateral epicondyle and sulcus on medial epicondyle and passing a k wire through it. Confirmation is done under image intensifier C arm with epicondylar view in Non navigated knees. Postoperative TEA was assessed by taking CT scan, measuring condylar twist angle and posterior condylar angle (PCA).
Results:
The mean PCA was around 4° with TEA as reference in Navigated and 6° in Non navigated knees and only 7% patients required an additional lateral release of which 2% patient had preop patellar maltracking. No postoperative patellar maltracking was seen. Anterior knee pain was present in 10% patients. No postop infection is noted. Alignment ranging from 4° to 8° external rotation.
Conclusion:
Navigation is most accurate measure for TEA as reference, as compared to non navigated TKA, which can lead to excessive external rotation especially in severely deformed knees.