Full Access

General Orthopaedics

Ct-Based Patient-Specific Instrumentation Is Accurate for Tka: A Single-Surgeon, Prospective Trial

International Society for Technology in Arthroplasty (ISTA)



Abstract

The present IRB approved study evaluates the early results of 100 TKAs using CT-based Patient-Specific Instrumentation (PSI) (MyKnee®, Medacta International, SA, Castel San Pietro, Switzerland). For this technique, a CT scan of the lower extremity is obtained, and from these images, the knee is reconstructed 3-dimensionally. Surgical and implant-size planning are performed according to surgeon preference, with the goal to create a neutral mechanical axis. Once planned and approved, the blocks are made [Fig. 1].

Outcomes measured for the present study include surgical factors such as Tourniquet Time (TT) as a measure of surgical efficiency, the actual intraoperative bony resection thicknesses to be compared to the planned resections from the CT scan, and complication data. Furthermore, pre- and post-operative long standing alignment and Knee Society Scores (KSS) were obtained.

During surgery, the PSI cutting block is registered on the femur first and secured with smooth pins. No osteophytes are removed as the blocks use the positive topography of the osteophytes for registration. The distal femoral resection is performed directly through the block. An appropriate sized 4-in-1 block is placed and the remaining resections are performed. The tibial resection block is registered and resection performed. Final bone preparation, patella resurfacing, and trialing is performed as is standard to all surgical techniques.

There were 50 Left and 50 Right TKA's performed in 61 females and 39 males. All patients had diagnosis of osteoarthritis. The average BMI was 31.1 and average age was 64.5 (range 41–90). 79 patients had pre-operative varus deformities with Hip Knee Angle (HKA) average of 174.7° (range 167°–179.5°). 19 patients had pre-operative valgus deformities averaging 184.4° (range 180.5°–190°). Three patients were neutral.

Average TT was 31.2 minutes (range 21–51 minutes). With regard to the bony resections, the actual vs. planned resections for the distal medial femoral resection was 8.7 mm vs. 8.9 mm respectively. Further actual vs. planned femoral resections include distal lateral 7.2 vs. 6.7 mm; posterior medial 8.3 vs. 8.9 mm; and posterior lateral 6.2 vs. 6.8 mm. The actual vs. planned tibial resections recorded include medial 6.4 vs. 6.3 mm and lateral 8.3 vs. 8.2. The planned vs. actual bony cuts are strongly correlated, and highly predictive for all 6 measured cuts (p=<.001) [Fig. 3]. No intraoperative complications occurred.

Average KSS improved from 45.9 to 81.4, and KSS Function Score improved from 57.7 to 73.5 at 6 weeks postoperative visit. There were no thromboembolic complications. Two patients had a post-operative infection requiring surgical intervention.

Post-operative alignment was 179.36° (range 175°–186°) for all patients. Alignment was neutral, within 3° in 95.9% of patients. There were only 4 outliers with maximal post-operative angulation of 6° [Fig. 2].

In conclusion, these early results demonstrate efficacy of CT-based PSI for TKA. The surgery can be performed efficiently, accurately, and safely. Furthermore, excellent short term clinical and radiographic results can be achieved.


*Email: