Abstract
Introduction:
Acetabular revision Jumbo cups are used in revision hip surgeries to allow for large bone to implant contact and stability. However, jumbo cups may also result in hip center elevation and instability. They may also protrude through anterior wall leading to ilopsoas tendinitis.
Methods:
The study was conducted using two methods:
Computer simulation study
265 pelvic CT scans consisting of 158 males and 107 females were converted to virtual 3-dimensional bones. The average native acetabular diameter was 52.0 mm, SD = 4.0 mm (males in 52.4 mm, SD = 2.8 mm and 46.4 mm, SD = 2.6 mm in females). Images were analyzed by custom CT analytical software (SOMA™ V.3.2)1 and over-sized reaming was simulated. Four distinct points, located in and around the acetabular margins, were used to determine the reamer sphere. Points 1, 2, 3 were located at the inferior and inferior-medial acetabular margins, and Point 4 was located superiorly and posteriorly in the acetabulum to simulate a bony defect in this location, Point 4 was placed at 10%, 20%, 30%, 40%, 50% and 60% of the distance from the superior – posterior margin of the acetabular rim to the sciatic notch to simulate bony defects of increasing size. (Figure 1)
Radiographical study
Retrospective chart review of patient records for all cementless acetabular revisions utilizing jumbo cups between January 1, 1998 and March 30, 2012 at UCFS (98 patients with 57 men, 41 women). Jumbo cups: ≥66 mm in males; <62 mm in females. Reaming was directed inferiorly to the level of the obturator foramen to place the inferior edge of the jumbo cup at the inferior acetabulum. To determine the vertical position of the hip center, a circle was first made around both the jumbo and the contralateral acetabular surfaces using Phillips iSite PACS software. The center of this circle was assumed to correspond to the “hip center”. The height of the hip center was estimated by measuring the height of a perpendicular line arising from the interteardrop line (TL) and ending at the hip center
Results:
The computer simulation and radiographic analysis deomonstrated similar results. The computer simulation predicted that the hip center shifted superiorly and anteriorly as the reamer size increased. The hip center shifted 0.27 mm superiorly and 0.02 mm anteriorly for every millimeter in diameter increased for the reaming. (Figure 2) Anterior column bone removal was increased 0.86 mm for every 1 mm of reamer size increase. (Figure 3)
Results of radiographical study is shown in Table bellow:
Discussion:
Use of a jumbo cup in revision THA results in elevation of the hip center. Therefore a longer femoral head may be needed to compensate for hip center elevation when a jumbo cup is used. Reaming for a jumbo cup can also result in loss of anterior bone stock and protrusion of the cup anteriorly which may cause iliopsoas tendonitis.