Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Oncology

KNEE ARTHRODESIS WITH A MODULAR PROSTHESIS SYSTEM: RESULTS AND COMPLICATIONS IN PRIMARY AND REVISION IMPLANTS

The European Musculo-Skeletal Oncology Society (EMSOS)



Abstract

Aim

To evaluate outcome and complications of knee arthrodesis with a modular prosthetic system (MUTARS(r) Implantcast), as primary and revision implants in musculoskeletal oncology.

Method

Between 1975 and 2009, 24 prostheses were used for knee arthrodesis. Nineteen in oncologic cases: 6 osteosarcomas, chondrosarcoma, synovial sarcoma and metastatic carcinoma 3 each, 2 pigmented villonodular synovitis (PVNS), malignant fibrous hystiocitoma and giant cell tumour 1 each. Patients were grouped into: A) primary implants, B) revision implants. Group A included 9 patients: 8 arthrodeses after extra-articular resection with major soft tissue removal, 1 after primary resection following multiple excisions of locally recurrent PVNS. Group B included 15 patients: 12 arthrodeses for infection (5 infected TKAs, 7 infected megaprostheses), 2 for failures of temporary arthodesis with Kuntscher nail and cement, 1 for recurrent chondrosarcoma in previous arthrodesis.

Results

Oncologic outcome ata mean follow-up of 6 years (ranging 1 to 26), showed 13 NED (68.4%), 2 NED after treatment of relapse (10.5%), 1 alive with metastases (5.3%), 2 dead with disease (10.5%) and one dead of other disease (5.3%).

Complications causing failure were observed in 12 patients (50%): 11 infections at mean of 14 months (6 in arthrodeses as revision for previous infections, 5 in group A), 1 femoral stem breakage at 4.8 years (in group B). Treatment of infections was: amputation in 6, “one stage” in 1, “two stage” with new arthrodesis in 4 (1 subsequently amputated for recurrent infection). The breakage was revised, had further traumatic breakage at 2 years and a second revision.

Conclusion

Arthrodesis with modular prosthetic system is indicated after major extra-articular resection or in revisions of severely failed previous reconstructions. High infection rate should be prevented with good soft tissue coverage, by flaps if needed.