Knee

TIBIAL TUNNEL PLACEMENT IN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION: A COMPARISON OF INTRA-OPERATIVE FLUOROSCOPY AND POST OPERATIVE RADIOGRAPHS

British Association for Surgery of the Knee (BASK)



Abstract

The outcome following arthroscopic anterior cruciate (ACL) reconstruction is dependant on a combination of surgical and non-surgical factors. Technical error is the commonest cause for graft failure, with poor tunnel placement accounting for over 80% of those errors.

A routine audit of femoral and tibial tunnel positions following single bundle hamstring arthroscopic ACL reconstruction identified apparent inconsistent positioning of the tibial tunnel in the sagittal plane. Intra-operative fluoroscopy was therefore introduced (when available) to verify tibial guide wire position prior to tunnel reaming. This paper reports a comparison of tibial interference screw position measured on post-operative radiographs with known tunnel position as shown on intra-operative fluoroscopic images in 20 patients undergoing routine primary ACL reconstruction between January and June 2009.

Surgery took a mean of 5 minutes longer when intra-operative fluoroscopy was used. In 3/20 patients, fluoroscopy led to re-positioning of the tibial guide wire prior to tunnel reaming. The mean tibial tunnel position as indicated by the tunnel reamer was 41 +/− 2.7 % of the total plateau depth (range 37% to 47%). The mean position projected from the tibial screw on post operative radiographs was 46 +/− 9.2% (range 38% to 76%). A paired t-test showed a significant difference (p = 0.022) between true tunnel position and tibial screw position. 6/20 patients had post operative screw positions that were > 5% more posterior than the known position of the tibial tunnel.

The position of the tunnel should be measured at its mid-point where this is evident. On most early radiographic images, the margins of the tunnel are not clear and therefore a line projected from the centre of the screw is used. This audit demonstrates the potential inaccuracy associated with this.