COMPARISON OF THE CT-BASED MICROMOTION ANALYSIS METHOD VERSUS MARKER-BASED RADIOSTEREOMETRIC ANALYSIS IN MEASURING FEMORAL STEM MIGRATION

The International Hip Society (IHS) 2024 Closed Meeting, Athens, Greece, 15–18 May 2024.



Abstract

Early micromotion of hip implants measured with radiostereometric analysis (RSA) is a predictor for late aseptic loosening. Computed Tomography Radiostereometric Analysis (CT-RSA) can be used to determine implant micro-movements using low-dose CT scans. CT-RSA enables a non-invasive measurement of implants. We evaluated the precision of CT-RSA in measuring early stem migration. Standard marker-based RSA was used as reference. We hypothesised that CT-RSA can be used as an alternative to RSA in assessing implant micromotions.

We included 31 patients undergoing Total Hip Arthroplasty (THA). Distal femoral stem migration at 1 year was measured with both RSA and CT-RSA. Comparison of the two methods was performed with paired-analysis and Bland-Altman plots. Furthermore, the inter- and intraobserver reliability of the CT-RSA method was evaluated.

No statistical difference was found between RSA and CTMA measurements. The Bland-Altman plots showed good agreement between marker-based RSA and CT-RSA. The intra- and interobserver reliability of the CT-RSA method was found to be excellent (≥0.992).

CT-RSA is comparable to marker-based RSA in measuring distal femoral stem migration. CTMA can be used as an alternative method to detect early implant migration.