Knee

LONG-TERM BIOMECHANICAL OUTCOMES OF A HIGH FLEXION DESIGN TOTAL KNEE ARTHROPLASTY

The British Association for Surgery of the Knee (BASK) May 2023 Meeting, London, England, 16–17 May 2023.



Abstract

Abstract

Introduction

This study compared biomechanical and functional parameters of a total knee arthroplasty (TKA) implant (Cemented Zimmer Hi-Flex) against healthy older adults to determine whether knee biomechanics was restored in this patient population.

Methodology

Patients with a primary TKA and healthy adults >55 years old with no musculoskeletal deficits or arthritis participated. Bilateral knee range of motion (RoM) was assessed with a goniometer, then gait patterns were analysed with a 3D motion-capture system. An arthrometer then quantified anterior-posterior laxity of each knee. Statistical analyses were performed in SPSS (α=0.05; required sample size: n=21 per group).

Results

25 knees were replaced in 21 patients. Nine presented with fixed flexion deformities (FFD) (13.3±5.6°). FFDs were abolished intraoperatively, and the average flexion increased from 124.8±9.1° to 130.9±5.8°. At 9.6±3.2 years postoperatively, the patients achieved poorer RoM than healthy controls (n=23); p<0.0001. These differences were due to limited flexion in the knee. Patients also failed to achieve the same degree of flexion as controls bilaterally during gait. No differences were observed during mid-flexion; a state that has been associated with instability (p=0.614). There were no differences between groups in knee laxity.

Conclusion

Patients in this study had similar gait patterns to healthy older adults during mid-flexion, and were no more likely to exhibit anterior-posterior translation of the knee >7mm; a known risk factor of instability. However, the flexion range was poorer. This led to bilateral pathological knee flexion patterns during gait. Further research should identify the cause of these limitations.