Full Access

Knee

ROSA VERSUS MAKO: COMPARING SURGICAL ACCURACY IN ROBOTIC TOTAL KNEE ARTHROPLASTY

The British Association for Surgery of the Knee (BASK) May 2023 Meeting, London, England, 16–17 May 2023.



Abstract

Abstract

Introduction

There is increasing adoption of robotic surgical technology in Total Knee Arthroplasty - The ROSA® knee system can be used in either image-based mode (using pre-operative calibrated radiographs) or imageless modes (using intra-operative bony registration). The Mako knee system is an image-based system (using a pre-operative CT scan). This study aimed to compare surgical accuracy between the ROSA and Mako systems with specific reference to Joint Line Height, Patella Height and Posterior Condylar Offset.

Methodology

This was a retrospective review of a prospectively-maintained database of the initial 100 consecutive ROSA TKAs and the initial 50 consecutive Mako TKAs performed by two high volume surgeons. To determine the accuracy of component positioning, the immediate post-operative radiograph was reviewed and compared with the immediate pre-operative radiograph. Patella height was assessed using the Insall-Salvati ratio.

Results

There was no significant difference between ROSA TKA and Mako TKA with regards to restoration of joint line height, ROSA mean 0.2mm versus Mako mean 0.3mm (p<0.05), posterior condylar offset, ROSA mean 0.16mm versus Mako mean 0.3mm (p<0.05), and patella height, ROSA mean 0.02 versus Mako mean 0.03 (p<0.05)

Conclusion

This study is the first study to compare the accuracy of the ROSA and MAKO knee systems in total knee arthroplasty. Both systems are highly accurate in restoring native posterior condylar offset, joint line height, and patella height in TKA with no significant difference demonstrated between the two robotic systems.