Knee

ROSA ROBOTIC ARM SYSTEM RELIABLY RESTORES JOINT LINE HEIGHT, PATELLA HEIGHT, AND POSTERIOR CONDYLAR OFFSET IN TOTAL KNEE ARTHROPLASTY

The British Association for Surgery of the Knee (BASK) May 2023 Meeting, London, England, 16–17 May 2023.



Abstract

Abstract

Introduction

There is growing interest in the use of robotic Total Knee Arthroplasty (TKA) to improve accuracy of component positioning. This is the first study to investigate the radiological accuracy of implant component position using the ROSA® knee system with specific reference to Joint Line Height, Tibial Slope, Patella Height and Posterior Condylar Offset. As secondary aims we compared accuracy between image-based and imageless navigation, and between implant designs (Persona versus Vanguard TKA).

Methodology

This was a retrospective review of a prospectively-maintained database of the initial 100 consecutive TKAs performed by a high volume surgeon using the ROSA® knee system. To determine the accuracy of component positioning, the immediate post-operative radiograph was reviewed and compared with the immediate pre-operative radiograph with regards to Joint Line Height, Tibial Slope, Patella Height (using the Insall-Salvati ratio) and Posterior Condylar Offset.

Results

Mean age of patients undergoing ROSA TKA was 70 years (range, 55 to 95 years). Mean difference in joint line height between pre and post-operative radiographs was 0.2mm (range −1.5 to +1.8mm, p<0.05), posterior condylar offset mean change 0.16mm (range −1.4 to +1.3mm, p<0.05), tibial slope mean change 0.1 degrees (p<0.05) and patella height mean change 0.02 (range −0.1 to +0.1 p<0.05). No significant differences were found between imageless and image-based groups, or between implant designs (Persona versus Vanguard).

Conclusion

This study validates the use of the ROSA® knee system in accurately restoring Joint Line Height, Patella Height and Posterior Condylar Offset.