THREE-YEAR AND FIVE-YEAR REVISION RATES OF PRIMARY TOTAL HIP ARTHROPLASTY ARE POOR PREDICTORS OF 15-YEAR REVISION RATES: A NATIONAL JOINT REGISTRY ANALYSIS

The British Hip Society (BHS) Meeting, Edinburgh, Scotland, 8–10 March 2023.



Abstract

Total hip arthroplasty has been constantly evolving with technological improvements to achieve the best survival rates. Although the new implants are under closer surveillance through processes such as Beyond Compliance, orthopaedic surgeons generally tend to look out for the latest implants with good short-term results and hope for better long-term results for these. We questioned whether such an assumption or bias is valid.

We analysed the data of Kaplan-Meier estimates of cumulative revisions of primary hip replacement by fixation, stem/cup brand and bearing combinations from the NJR 19th Annual Report published in September 2022. We performed a univariate linear regression analysis to predict the 10- and 15-year revision rates for these different hip implant combinations from the 3- and 5-year revision rates.

Thirty-seven implant combinations had their 15-year revision rates reported and 67 had the 10-year revision rates. The correlation co-efficients were 0.43 and 0.58 for the 3-year and 5-year revision rates against 15-year revision rates. Only 17% of the variance in 15-year revision rates could be predicted by a linear regression model from the 3-year revision rate and 32% from the 5-year revision rate. Corresponding values for the 10-year revision rates were 46% and 67%.

95% prediction intervals for the 15-year revision rate were +/− 3.1% from the 3-year revision rate and +/− 2.8% from the 5-year revision rate. Corresponding values for the 10-year revision rates were +/− 1.3% and +/− 1%.

19 of 37 implant combinations showed 15-year revision rate of more than 4%. Average 3-year and 5-year revision rates for this cohort was 1.0% and 1.42% compared to 1.4% and 1.9% for the rest and the difference was statistically significant.

Although average early revision rates showed small but significant difference between the groups with lower and higher 15-year revision rates, the prediction intervals for 15-year revision rates for individual hips based on their 3-year and 5-year revision rates are very wide. Three- and 5-year revision rates for primary total hip replacements are poor predictors of 15-year revision rates.