Abstract
Abstract
Introduction
Cementless fixation of Oxford Unicompartmental Knee Replacements (UKRs) is an alternative to cemented fixation, however, it is unknown whether cementless fixation is as good long-term. This study aimed to compare primary and long-term fixation of cemented and cementless Oxford UKRs using radiostereometric analysis (RSA).
Methodology
Twenty-nine patients were randomised to receive cemented or cementless Oxford UKRs and followed for ten years. Differences in primary fixation and long-term fixation of the tibial components (inferred from 0/3/6-month and 6-month/1-year/2-year/5-year/10-year migration, respectively) were analysed using RSA and radiolucencies were assessed on radiographs. Migration rates were determined by linear regression and clinical outcomes measured using the Oxford Knee Score (OKS).
Results
Preliminary analysis of Maximum Total Point Motion (MTPM) indicated cementless tibial components undergo significantly more migration than cemented components during the first 6 months (1.6mm/year, SD=0.92 versus 1.3mm/year, SD=1.1, p<0.001). Cementless migration was predominantly subsidence inferiorly (Mean=0.51mm/year, SD=0.29, p<0.001) and posteriorly (0.13mm/year, SD=0.21, p=0.03). Contrastingly, from 6 months to 10 years cemented components migrated significantly (MTPM=0.039mm/year, SD=0.11, p=0.04) whereas cementless components did not (MTPM=0.002mm/year, SD=0.02, p=0.744). Radiolucent lines occurred more frequently below cemented (10/13) than cementless (4/16) tibial components, but radiolucencies did not correlate with differences in migration or OKS. There was no significant difference in OKS between cemented and cementless.
Conclusion
These results suggest that cementless tibial components migrate more than cemented before achieving primary fixation. However, long-term fixation of cementless tibial components appears to be as good, if not better, than cemented with the benefit of fewer radiolucent lines.