Abstract
Abstract
Background
Little scientific evidence is available regarding the effect of knee joint line obliquity (JLO).
Methods
10 fresh-frozen human cadaveric knees were axially loaded to 1500 N in a materials testing machine with the joint line tilted 0, 4, 8, and 12 degrees varus and valgus, at 0, and 20 degrees of knee flexion. The mechanical compression axis was aligned to the centre of the tibial plateau. Contact pressures / areas were recorded by sensors inserted between the tibia and femur below the menisci. Changes in relative femoral and tibial position in the coronal plane were obtained by an optical tracking system.
Results
medial and lateral JLO caused significant tibiofemoral subluxation and pressure distribution changes. Medial (varus) JLO caused the femur to sublux medially down the coronal slope of the tibial plateau, and vice versa for lateral (valgus) downslopes (P=0.01). Areas of peak pressure moved 12 mm and 8 mm across the medial and lateral condyles, onto the ‘downhill’ meniscus and the ‘uphill’ tibial spine. Changes in JLO had only small effects on maximum contact pressures.
Conclusion
A change of JLO during load bearing caused significant mediolateral tibiofemoral subluxation. The femur slid down the slope of the tibial plateau to abut the tibial eminence and also to rest on the downhill meniscus.
Clinical Relevance
These results provide important information for understanding the consequences of creating coronal JLO and for clinical practice in terms of osteotomy planning regarding the effect on JLO.