Abstract
Abstract
Introduction
Higher than expected rates of tibial loosening with the ATTUNE® total knee arthroplasty (TKA) implant has been reported. Component loosening can be associated with the development of radiolucency lines (RLL) and our study aim was to systematically assess the reported rates of these.
Methodology
A systematic search was undertaken using the Cochrane methodology in four online databases. Identified studies were assessed and screened against predetermined inclusion criteria. Meta-analysis was conducted using a random-effects model.
Results
Nine studies (n=2,727 TKAs) from 6,590 titles met the inclusion criteria: 1 Randomised Controlled Trial (n=74), 1 prospective cohort (n=200), 4 retrospective cohort (n=1,639), and 3 case-series (n=814). All used the 2013 ATTUNE® design. In meta-analysis: 8 studies (n=1,440 ATTUNE TKAs) reported an overall prevalence of 11% (95%CI: 6.4-18.3%) for medial tibia RLL; 7 studies (n=940) a 12.3% (95%CI: 4.0-32%) rate of any tibia RLL.; 5 studies (n=736) femoral RLL in 11% (95%CI: 7.2-106.5%) and 7 studies (n=896) any RLL in 20.7% (95%CI: 13.4-30.6%). Meta-analysis of 4 studies (n=1,036) comparing the ATTUNE® with another implant (PFC Sigma®, LCS®, or PERSONA®) showed a higher risk of medial tibia RLL (OR: 2.538; 95%CI: 1.397-4.611, P=0.002) and any RLL (OR: 2.725; 95%CI: 1.302-5.703, P=0.008) in the ATTUNE® group.
Conclusions
The 2013 ATTUNE® TKA system is associated with high rates of radiolucency around the tibial and femoral components. Comparative studies suggesting these rates are more than double those of other systems. Radiolucency may be a herald of component loosening, therefore, close surveillance of this implant is recommended.