EXETER ANALYSIS OF ROBOTIC TOTAL HIP ARTHROPLASTY (EARTH): INITIAL RESULTS OF THE FIRST 40 CASES PERFORMED IN AN NHS HOSPITAL

The British Hip Society (BHS) Meeting, Bournemouth, England, 2–4 March 2022.



Abstract

The MAKO Robotic arm is a haptic robotic system that can be used to optimise performance during total hip arthroplasty (THA). We present the outcome of the first 40 robotic cases performed in an NHS foundation trust along with the technique of performing robotic THA in our unit.

Forty consecutive patients undergoing robotic THA (rTHA) were compared to a case matched group of patients undergoing manual THA (m-THA). 2:1 blinded case matching was performed for age, sex, implants used (Trident uncemented socket and cemented Exeter stem, Stryker Mahwah, NJ, US) and surgeon grade. Comparisons were made for radiological positioning of implants, including leg length assessment, and patient reported functional outcome (PROMS). Pre- and post-operative radiographs were independently analysed by 2 authors.

All patients underwent THA for a primary diagnosis of osteoarthritis. No significant difference between groups was identified for post-operative leg length discrepancy (LLD) although pre-operatively a significantly higher LLD was highlighted on the MAKO group, likely due to patient selection. Significantly lower post-operative socket version was identified in the MAKO cohort although no difference in post-operative cup inclination was noted. However, there was significantly larger variance in post-op LLD (p=0.024), cup version (p=0.004) and inclination (p=0.05) between groups indicating r-THA was significantly less variable (Levene's test for homogeneity of variance). There was no significant difference in the number of cases outside of Lewinnek's ‘safe’ zone for inclination (p=0.469), however, there were significantly more cases outside Lewinnek's ‘safe’ zone for version (12.5% vs 40.3%, p=0.002) in the m-THA group.

We report the commencement of performance of MAKO robotic THA in an NHS institution. No problems with surgery were reported during our learning curve. Robotic THA cases had less variability in terms of implant positioning suggesting that the MAKO robot allows more accurate, less variable implant positioning with fewer outliers. Longer term follow-up of more cases is needed to identify whether this improved implant positioning has an effect on outcomes, but the initial results seem promising.