SPINOPELVIC MOBILITY AFFECTS ACCURACY OF ACETABULAR ANTEVERSION ON CROSS-TABLE LATERAL RADIOGRAPHS

The Hip Society (THS) 2020 Members Meeting, held online, 1–2 October 2020.



Abstract

Introduction

Cross table lateral (CTL) radiographs are commonly used to measure acetabular component anteversion after total hip arthroplasty (THA). CTL measurements may differ by >10 degrees from CT scan measurements, but the reasons for this discrepancy are poorly understood. We compare anteversion measurements made on CTL radiographs and CT scans to identify spinopelvic parameters predictive of inaccuracy.

Methods

THA patients (n=47) with preoperative spinopelvic radiographic analysis and postoperative CT scans were retrospectively reviewed. Acetabular component anteversion was measured on post-operative CTL radiographs, and CT scans using 3D reconstructions of the pelvis. Patients were grouped by error (CTL-CT)>10° (n=11) or <10° (n=36), and spinopelvic mobility parameters were compared using t-tests. Correlation between error and mobility parameters was assessed with Pearson coefficient.

Results

Patients with CTL error >10° (range 10–14) had stiffer lumbar spines with less lumbar flexion (38° vs 47°, p=0.03), greater sagittal imbalance measured by pelvic incidence-lumbar lordosis mismatch (6° vs −2°, p=0.04), more pelvic extension when seated (pelvic tilt −10° vs −2°, p=0.05), and greater change in pelvic tilt between supine and seated positions (13° vs 4°, p=0.04). The error of CTL measurements showed a positive correlation with increased CTL anteversion (r=0.5, p=0.001), standing lordosis (r=0.23, p=0.05), seated lordosis (r=0.4, p=0.01) and pelvic tilt change between supine and step-up positions (r=0.34, p=0.01).

Discussion

Differences in spinopelvic mobility patterns may explain the variable accuracy of acetabular anteversion measurements on CTL radiographs. Patients with stiff spines and increased compensatory pelvic motion have less accurate measurements on CTL radiographs. Flexion of the contralateral hip is required to obtain clear CTL radiographs. In patients with a stiff lumbar spine, this movement may extend the pelvis and increase anteversion of the acetabulum on CTL views. Reliable analysis of acetabular component anteversion in this patient population may require advanced imaging with a CT scan.