Hip

MID-TERM SURVIVORSHIP OF A CONSTRAINED ACETABULAR DEVICE WITH INCREASED RANGE OF MOTION TO HELP PREVENT DISLOCATION

The Hip Society (THS) 2019 Summer Meeting, Kohler, WI, USA, 25–27 September 2019.



Abstract

Background

Recurrent instability and dislocation after total hip arthroplasty are difficult complications. In certain cases, a constrained acetabular device can address these issues. The purpose of this study is to report the midterm outcomes and survivorship of a single constrained liner device.

Methods

A retrospective review as performed on all procedures (except first stage exchange for infection) in which a FreedomĀ® Constrained (Zimmer Biomet, Warsaw, IN) liner was used between December 2003 to November 2016. Patients with 2-year minimum follow-up or failure were included, yielding a cohort of 177 patients. Procedures were 130 revisions, 40 reimplantations following infection eradication, and 7 complex primaries. The constrained mechanism was implanted in 46 hips (26%) to treat active instability and 131 hips (74%) for increased risk of instability or intraoperative instability. Patients had on average 3.4 previous surgeries.

Results

With an average 7.1-year follow-up, 11 hips dislocated (6.2%), and 13 hips (7.3%) were revised for aseptic loosening of the acetabular component, resulting in an overall constrained mechanism aseptic or mechanical failure rate of 14.1%. Nineteen hips (10.7%) failed from infection with 58% of these having had a previous infection. Patients with active instability had significantly higher failure for dislocation than patients who were at-risk (15.2% vs. 3%, p=0.01). All-cause survival rate at 7 years was 74.8%%, aseptic survival was 83.6%, and survival for instability was 91.8%.

Conclusion

Revision for instability remains challenging as many patients have had numerous previous surgeries and at-risk anatomy. Constrained inserts are one option to manage instability, but a high rate of recurrence still occurs.