FEMORAL HEAD PENETRATION AND CLINICAL RESULTS WITH DUAL MOBILITY VITAMIN E-INFUSED POLYETHYLENE IN PRIMARY TOTAL HIP ARTHROPLASTY: EARLY RESULTS

The Hip Society (THS) 2019 Summer Meeting, Kohler, WI, USA, 25–27 September 2019.



Abstract

Introduction

Dual-mobility (DM) articulations may be useful for patients at increased risk for instability in primary and revision THA. While DM articulations are becoming increasingly popular, its routine use in primary THA is more uncertain. Even less is known about femoral head penetration in DM designs manufactured with highly cross-linked polyethylene infused with Vitamin E (E-HXLPE). The purpose of this study was to evaluate the early clinical results and femoral head penetration rates of primary THA implanted with DM E-HXLPE.

Methods

Between 2012 and 2017, 105 primary DM THAs were performed using a one-piece acetabular shell, 28mm ceramic head, coupled with an E-HXLPE outer bearing via a standard posterior approach. Three patients refused follow-up after six months. 102 hips (92 patients) were available for review. The diagnosis was 99% OA. Average age was 65.7 years (33–90 years). 56% of patients were female. The most common femoral head size was 50mm (range, 44–60mm). The average thickness of the E-HXLPE outer bearing was 22.7mm (range, 16–32mm). Patients were followed at two months (baseline radiograph), six months, one, three, five, and seven years. Harris hip scores (HHS), UCLA activity score, and femoral head penetration (Martell method) were obtained at each visit beyond two months. Follow-up averaged 3 years (range, 1–7 years).

Results

Average HHS improved from 43 to 95 points (50–100) at final follow-up. Similarly, average Harris pain scores improved from 10 to 42 points (20–44) with 78% of patients pain free and one patient, each, reporting groin and thigh pain (1%). Average UCLA activity scores was 8.1 (range, 5–10). There were no dislocations, revisions, or loose cups. The average femoral head penetration (including bedding-in) was 0.25mm/yr (s.d. 0.2mm/yr) at seven years.

Discussion/Conclusion

The theoretical benefits of DM designs in diminishing wear include a smaller inner head, lower frictional torque with motion, less micro-separation, and less wear with impingement. The early clinical results of primary DM THA are promising. Although the early femoral head penetration using DM E-HXLPE appears to be less than other DM designs, it was greater than that of conventional THA using HXLPE and E-HXLPE inserts. The additional outer bearing surface, while affording additional stability, may actually enhance polyethylene creep and wear.