Abstract
Manufacturers of reverse shoulder arthroplasty (RSA) implants have recently designed innovative implants to optimise performance in rotator cuff deficient shoulders. These advancements are not without tradeoff and can have negative biomechanical effects. The objective of this study was to develop an integrated FEA kinematic model to compare the muscle forces and joint reaction force (JRF) of 3 different RSA designs.
A kinematic model of a normal shoulder joint was adapted from the Delft model and integrated with the OpenSim shoulder model. Static optimisations then allowed for calculation of the individual muscle forces, moment arms and JRF relative to net joint moments. Three dimensional computer models of humeral lateralised design (HLD), glenoid lateral design (GLD), and Grammont design (GD) RSA were integrated and parametric studies were performed.
Overall there were decreases in deltoid and rotator cuff muscle forces for all 3 RSA designs. These decreases were greatest in the middle deltoid of the HLD model for abduction and flexion and in the rotator cuff muscles under both internal and external rotation. The joint reactive forces in abduction and flexion decreased similarly for all RSA designs compared to the normal shoulder model, with the greatest decrease seen in the HLD model.
These findings demonstrate that the design characteristics implicit in these modified RSA prostheses result in kinematic differences most prominently seen in the deltoid muscle and overall joint reactive forces. Further research utilising this novel integrated model can help guide continued optimisation of RSA design and clinical outcomes.