Abstract
As previous meta-analyses on the alignment outcomes of Computer-assisted orthopaedic surgery (CAOS) did not differentiate between CAOS systems, limited information is available on the accuracy of a specific CAOS system based on clinical cases. This study assessed the accuracy and precision of achieving surgical goals in approximately 7000 cases using a specific contemporary CAOS system.
Alignment parameters were extracted from the technical logs of 6888 TKA surgeries performed between October 2012 and January 2017 using a contemporary CAOS system. The following surgical parameters were investigated: 1) planned resection defined by the surgeon prior to the bone cuts; 2) Checked resection defined as digitalisation of the bony cuts. Deviations in alignment between planned and checked resections were evaluated, with acceptable resections defined as no more than 3° of resection deviations.
For the tibial resection, deviations in tibial varus/valgus angle and posterior tibial slope were 0.06 ± 0.94° and −0.09 ± 1.73°, respectively. For the femoral resection, deviations in femoral varus/valgus angle amd femoral flexion were 0.00 ± 0.97° and −0.17 ± 1.44°, respectively. High percentages of the resections were found to be acceptable (>94% of the cases).
The CAOS system investigated was shown to provide accurate and precise intra-operative assistance to the surgeon in achieving targeted resections. The study summarised a large number of cases spanning the application history of the specific CAOS system, including both experienced users and new adopters of the technology. The data provided a complete clinical relevant evaluation demonstrating its high accuracy and precision in resection alignment.