Abstract
Acoustic emission is an uncommon but well-recognised phenomenon following total-hip arthroplasty using hard-on-hard bearing surfaces. The incidence of squeak has been reported between 1% – 10%. The squeak can be problematic enough to warrant revision surgery. Several theories have been proposed, but the cause of squeak remains unknown. Acoustic analysis shows squeak results from forced vibrations that may come from movement between the liner and shell. A potential cause for this movement is deformation of the shell during insertion.
6 cadaver hemipelvises were prepared to accept ace-tabular components. A shell was selected and pre-insertion the inner shape was measured using a profilometer. The shell was implanted and re-measured. 2x screws were then placed and the shells re-measured. The results were assessed for deformation.
Deformation of the shells occurred in 5 of the 6 hemi-pelvises following insertion. The hemipelvis of the non-deformed shell fractured during insertion. Following screw insertion no further shell deformation occurred.
The deformation was beyond the acceptable standards of a morse taper which may allow movement between components, and this may produce an acoustic emission. Further in-vitro testing is being conducted to see whether shell deformation allows movement producing an acoustic emission.