Full Access

033 DELAYED FOREIGN BODY REACTION TO ABSORBABLE IMPLANTS IN METACARPAL FRACTURE TREATMENT



Abstract

The aim of this study was to explore whether adverse reactions would occur during the material’s degradation period even at a later time point after surgery and whether these phenomena were clinically significant and would influence the final outcome.

12 unstable, displaced metacarpal fractures in 10 patients (7 males, 3 females; mean age 36.4 y, range 18–75 y) were treated with the Inion® OTPSTM Biodegradable Mini Plating System. 9 patients (10 fractures) were available for follow-up (mean 25.6 months, range 14 to 44 m). For patients without appearance of foreign body reaction the minimum follow-up time was 24 months

Patients were examined both radiologically to evaluate fracture healing, and clinically by completing the DASH-score and a visual analogue scale for pain assessment. Grip strength, finger strength and range of motion of metacarpo-phalangeal and interphalangeal joints were measured.

Fracture healing occurred uneventfully in all patients within six weeks. The most important complication was a foreign body reaction observed in 4 of our patients more than a year postoperatively. All were re-operated and had the materials removed. Histological examination confirmed the diagnosis of aseptic inflammation and foreign body reaction.

Although internal fixation of metacarpal fractures by using bioabsorbable implants is a satisfactory alternative fixation method, patients should be advised of this possible late complication and should be followed postoperatively for at least one and a half year, possibly longer.

Correspondence should be addressed to Anastasia C. Tilentzoglou MD, General Secretary of the Board of Directors of HAOST, 20 A. Fleming Str. (N.Filothei), Gr. 15123 Maroussi, Athens Greece. E-mail: info@eexot.gr