Abstract
Introduction: Potential problems resulting from metal-on-metal hip wear debris are well known. Previous studies have shown an association between high cup angles and raised metal ion levels, but a link to clinical outcome has yet to be established. We aim to show the relationship between high cup angles, raised metal ion levels and pain following hip resurfacing.
Methods: This analysis comprised the first 250 Depuy ASR hip resurfacings performed by a single surgeon (senior author) over a 3 yr period at an independent centre. Patients were followed up, examined and x-rayed at regular intervals. We measured Harris Hip Score (HHS) and pain. The acetabular cup inclination angles were measured from standardised pelvic radiographs. A subgroup of 80 patients had serum and whole blood chromium and cobalt measurements performed. This subgroup comprised patients with a similar sex split, age, activity level, follow-up and cup angle profile to the parent group.
Results: HHS improved from 51.6 pre-operatively to 94.6 post-operatively. Males had a significantly higher post-op HHS (97.1) compared with the females (91.0). 2.1% of male patients had pain compared with 8.7% of females. High cup angles were associated with pain in females. This relationship did not occur in males. In patients with cup angles of 48° and above there were no males with pain compared to 15.4% of females. Females under 48° had a HHS of 93.7 compared to 88.0 for those over 48°. When we analysed metal ion levels there was a similar relationship. Patients with cup angles of 48° and above had significantly higher serum and whole blood chromium and cobalt levels compared with lower cup angles. Three patients with cup angles over 50° developed ALVAL (aseptic lymphocyte dominated vasculitis associated lesion) based on histological findings. All were female.
Discussion: High cup angles are associated with increased pain and elevated metal ion levels. Women seem to be more intolerant of an higher cup angle than men. This may be a consequence of the smaller prostheses in females. Cups implanted lower than 48° had a better clinical outcome and less metal ion wear. Like other 4th generation designs, the ASR cup is not a complete hemisphere (unlike the BHR) and may be susceptible to edge loading at lower inclination angles than previously thought. We believe the current recommendation of 45° +/−5° is too high based on our findings. We recommend an inclination angle that does not exceed 48°.
Correspondence should be addressed to: EFORT Central Office, Technoparkstrasse 1, CH – 8005 Zürich, Switzerland. Email: office@efort.org