Abstract
The early results with highly cross-linked polyethylene have been encouraging and have increased the ability to use larger head diameters to improve the range of motion and decrease the dislocation rate, the commonest cause of early complications following total hip arthroplasty (THA). Wear rates with 32 mls heads have been satisfactory however there have been very few independent studies looking at early polyethylene wear in 36 mm heads. This study assessed the rate of polyethylene wear of a 36mm ceramic femoral head and a highly cross-linked polyethylene (X3 Stryker) liner in THA.
This prospective study reviewed 100 consecutive THAs in young patients (mean age 58 years) who had undergone THA with the same 36mm ceramic femoral head and highly cross linked polyethylene liner. All patients received the same femoral stem (ABG, Stryker) and acetabular cup (Trident, Stryker). Two surgeons performed all procedures. Patients were assessed radiologically immediately postoperatively, at 10 weeks and at one year. Validated computer software (Polyware) was used to assess both volumetric and linear wear.
At one year the mean two-dimensional linear wear rate was 0.51 mm/yr. Mean three-dimensional linear vector wear rate was 0.59 mls per year with a mean volumetric wear rate of 322.6 mms per three years. Cup size ranged from 52–62 mms and the correlation coefficient between cup size and three-dimensional linear wear rate was −0.100. The correlation coefficient between cup size and volumetric wear rate was −0.009 confirming no significant correlation between cup size and wear.
Larger size femoral heads are associated with a higher volumetric wear compared to linear wear rate when using conventional polyethylene. This study demonstrated much higher early linear wear rates compared to other studies using 28 and 32 mms heads. This higher rate may be associated with the creep phenomenon and early bedding-in in the early stages after a THA and although this is of concern these results should be interpreted with caution.
Correspondence should be addressed to Associate Professor N. Susan Stott, Orthopaedic Department, Starship Children’s Hospital, Private Bag 92024, Auckland, New Zealand.