HP8: PULSE LAVAGE AND HYDROGEN PEROXIDE FOR BONE PREPARATION PRIOR TO CEMENTING FEMORAL PROSTHESES



Abstract

There are several different ways of preparing the femoral canal prior to cementing a hip prosthesis. This study investigated the mechanical strength of the cement-bone interface of four different types of preparation determined by the maximum tensile force required to separate a cemented prosthesis from its cancellous bone origin.

Forty-eight fresh-frozen ox femora were prepared for hip arthroplasty, In a four-way comparison, groups of eleven femora were prepared by irrigation using

  1. syringe injected normal saline;

  2. hydrogen-peroxide soaked gauze;

  3. pulse-lavage brushing; and

  4. pulse-lavage brushing and hydrogen-peroxide soaked gauze combination.

Specimens were secured to a Material-test System (MTS), and the femoral implant pulled from the femur uni-axially at a rate of 5mm/min. The ‘pull-out strength’ was defined as the maximum tension recorded by the MTS during separation. Cement interdigitation was also inspected for each technique by microscopy of eight bone-implant transverse sections taken from prepared specimens.

Following an analysis of variance and pair-wise Fisher comparison, the average pull-out strength of the cemented prosthesis was significantly higher (P< 0.001) using pulse-lavage brushing (mean 8049.2 N), and pulse-lavage brushing in combination with hydrogen-peroxide soaked gauze (mean 8489.1 N), than with normal saline irrigation (mean 947.1 N) or hydrogen-peroxide soaked gauze preparation (mean 1832.6 N). Prosthesis pull-out strength following pulse-lavage brushing in combination with hydrogen-peroxide soaked gauze was not significantly different (P> 0.05) than preparing with pulse-lavage brushing alone. Low and high power microscopy of specimen transverse sections revealed the greatest levels of cement penetration in specimens prepared using pulse-lavage brushing.

This study demonstrated that one of the most effective preparations of the femoral canal for optimal mechanical fixation between cement and cancellous bone is pulse- lavage brushing. The use of hydrogen-peroxide soaked gauze in femoral canal preparation, either alone or in combination with pulse-lavage brushing, may not significantly improve prosthesis fixation.

The abstracts were prepared by David AF Morgan. Correspondence should be addressed to him at davidafmorgan@aoa.org.au