CORRECTION OF CUTTING ERRORS WITH NAVIGATION SYSTEM IN TOTAL KNEE REPLACEMENT



Abstract

Achieving precise component alignment of total knee arthroplasty produces good clinical outcome. However, the cutting errors between planed and final bone resection planes during the procedure of total knee arthroplasty were less evaluated. The aim of this study was to evaluate the cutting errors during total knee arthroplasty using the navigation system.

In a prospective series of 60 total knee replacements with image-free navigation system, the planed resection plane and final resection plane in frontal and sagittal planes were evaluated. The cutting errors standard deviations ranged from 1.01° to 1.21° in final frontal femoral and tibia plane and 1.23° in final sagittal femoral and tibia plane. The cutting errors showed only significant difference in the sagittal plane of femoral resection and only 9 cuts (4%) 3 of all plane and the maximal error was 4 in only 2 cases (0.8%).

Our results support to use the navigation system to adjust the cutting block and correct the cutting errors. This would lead to a more precise cut and result in better leg alignment and component orientation than the conventional TKR technique.