Abstract
The introduction of porous tantalum metal (Trabecular Metal; Zimmer, Warsaw, IN) for acetabular component fixation in total hip arthroplasty has shown optimum fixation qualities and “gap filling” effect. Recently, trabecular metal was introduced in tibial component for total knee prosthesis, however its effect on the bone mineral density (BMD) was not reported. The purpose of this study was to compare the BMD of proximal part of the tibia between trabecular metal and another cemented tibial component.
31 knees receiving trabecular metal tibial component and 33 knees receiving cemented tibial component (PFC Sigma RP, Depuy, Warsaw, IN) had dual energy x-ray absorptiometry (DEXA) scans at preoperatively and 3 weeks, 3, 6, 12, 18, 24 months post-operatively. To assess peri-prosthetic BMD, three regions of interest (ROI) were measured for each case. They were medial aspect (ROI 1), center aspect (ROI 2) and lateral aspect (ROI 3) of tibia. Average follow up period was 1.8 (range: 1.5 to 2) years. In both groups, BMD in tibia decrease postoperatively. Comparing postoperative decrease of BMD in lateral aspect of tibia (ROI 3) between both groups, it was significantly less in trabecular metal component (−0.09 g/cm2 +/−0.27) than cemented tibial component (−0.31 g/cm2 +/− 0.21) (p=0.0007).
We conclude that trabecular metal tibial component showed a favorable effect on BMD of proximal part of the tibia after total knee arthroplsaty.