Abstract
Purpose: To investigate whether cement mantle thickness influence early migration of the stem after impaction grafting
Methods: Twelve artificial femora were prepared to mimic cavitary defects. After compacting morselized bone into the cavities, Exeter stems were cemented in place. By using all combinations of three sizes tamps and stems (0, 1 and 2), we created cement mantles of 0, 1, 2, 3 and 4 mm thickness. Bones with stems were placed in a testing machine and loaded cyclically to 2,500 N while measuring stem migration. Statistical analysis was by regression analysis. Outcomes were stem subsidence and retroversion, predictors were mantle thickness, tamp size and stem size.
Results: Average stem subsidence after 2500 cycles when using size 1 tamp and stem (2 mm mantle) was 0.94 mm. Cement mantle thickness significantly influ-enced stem subsidence (r=0.68, p=0.015). For a 0 mm mantle, subsidence was 0.59 mm and for a 4 mm mantle it was 2.54 mm. Cement mantle thickness also signifi-cantly influenced stem retroversion (r=0.62, p=0.031). Cement mantle thickness was a better predictor than tamp or stem size.
Discussion: Concern exists that inadequate cement mantles may affect stability of impaction-grafted stems. In our study, larger difference between tamps and stems gave substantially more subsidence and rotation, whereas a smaller difference reduced them. Concerns over thin mantles may have been premature.