CEMENT RESTRICTOR FUNCTION BELOW THE FEMORAL ISTHMUS



Abstract

Introduction: Cement pressurisation is key to achieving good cement-bone interdigitation in THR. To obtain adequate pressurisation the medullary canal must be sealed distally using a cement restrictor. The cement restrictor must remain stable in the femoral canal.

Methods: Five different cement restrictors were evaluated, namely the Exeter Cement Plug, Biostop G, Hardinge, Rex CementStop and a preinjected cement plug. The restrictor was deployed in a sawbone that had been reamed to produce a distal flare, based on radiographic measurements. Low viscosity bone cement pressurised using a cement ram connected to a 10bar air supply. An electronic pressure valve increased the pressure in the cement. Cement pressure and cement restrictor displacement were continuously measured. The pressure valve and recording of measurements was controlled by a customised computer package.

Results: The Rex CementStop withstood the greatest pressures (mean 565.8kPa). This was a significantly greater pressure than any of the other cement restrictors (p< 0.001). Pre-injected cement plugs were able to resist the next highest pressures (mean 350.4kPa). They did not displace but leaked cement and were technically difficult to deliver in the distal femur. Cement restrictors that function well above the isthmus were ineffective (Biostop mean 118.7kPa) or could not be deployed below the isthmus (Exeter). The Hardinge recorded a mean 162.3kPa.

Discussion: During pre-operative templating it is important to consider where the cement restrictor will sit in the femur. When the cement restrictor is going to be deployed beyond the femoral isthmus, an alternate method of cement restriction may need to be used. Universal sized plugs (e.g. Hardinge) function poorly in this situation. Press-fit plugs such as Biostop and Exeter are severely compromised when inserted past the femoral isthmus. Pre-injected cement plugs are variable in efficacy. The expandable Rex CementStop reliably occluded the femur, allowing the highest pressures to be generated.