Abstract
Anatomic reduction and appropriate implant placement is essential for optimal treatment of intraarticular tibial plateau fractures. Standard intraoperative fluoroscopy provides limited visualization of the reduction and hardware placement compared with pre- or postoperative 3-D imaging modalities. As such, post-operative computer tomography (CT) has become a common procedure to evaluate the quality of the reduction and fixation. The Iso-C3D provides 3-D intraoperatively imaging to dynamically assess the surgical reduction and fixation at different anatomic regions. We report on our first 19 clinical tibial plateau fractures scanned intra-operatively with the Iso-C 3D.
Between January and November 2003, 19 intraarticular tibia plateau fractures were scanned intraoperatively with the Iso-C3D (Siemens, Germany). No formal selection criteria were utilised except for the presence of a tibial plateau fracture. Operative procedures included 14 cases of open reduction internal fixation and 5 cases of internal fixation with arthroscopic assisted reduction.
Imaging Technique: All patients were positioned on full-carbon tables for the operative procedure. After initial operative reduction and fixation, conventional two-dimensional fluoroscopic imaging was performed using standard AP and lateral projections. These images were evaluated by the operating surgeon; if the reduction and fixation was judged to be appropriate, Iso-C3D imaging was initiated
In 21% (n=4) of all cases an immediate revision of the operative procedure was performed after Iso-C3D imaging. These revisions were not deemed necessary with conventional fluoroscopy alone. In two cases, significant intra-articular incongruencies (greater than two millimetres) were noted. Additionally, in two cases, implant mal-position was detected. All patients had a postoperative CT scan. All CT scans confirmed the intraoperative Iso-C imaging, no further additional articular incongruencies or malpositioned implants were identified. When compared to conventional C-arm images, the Iso-C 3D scans demonstrated improved ability to identify the articular malreduction and implant mal-position in all cases.
We have demonstrated that the Iso-C3D provides reliable intraoperative evaluation of reduction and hardware placement compared to traditional CT scans for tibial plateau fractures. In addition, clinically relevant intra-operative information was gained with its use in this study. In four (21%) cases, the operative treatment was modified due to the use of the multiplanar imaging modality. On average, 10 minutes of additional operative time was required for the use of Iso-C3D scanning and the evaluation of the images. Further prospective clinical studies are needed to improve our findings.