Abstract
Purpose: Primary bone tumours situated at the distal end of the humerus are exceptional and observed in only 1% of the cases. Reconstruction at this level is often difficult due to the wide bone resection. Choosing a massive prosthesis is an interesting method allowing correct carcinological and functional results. We report a series of ten patients who underwent reconstruction of the lower end of the humerus between 1970 and 2001 with a massive prosthesis after tumour resection.
Material and methods: This series included four men and six women, mean age 51 years (15–76). Eight patients had primary bone tumours and two had secondary bone lesions. Mean follow-up was 79 months (9–372). The custom-made constrained hinge prosthesis composed of chromium-cobalt and titanium was cemented. Regular clinical and radiological follow-up data were recorded. The Henneking score and the Toronto Extremity Survival Score (TESS) were used to assess functional outcome.
Results: Three patients underwent revision for aseptic loosening of the humeral component at 48, 56 and 366 months with problem of polyethylene insert wear for two of them. There were no infctions or local recurrences and no secondary amputations. The posterior or anterolateral approach was used for humeral resection (mean 153 mm, 63–160 mm). Postoperatively, three patients developed transient palsy (one radial and two ulnar) which regressed. Four patients died from metastasis and had a satsifactory prosthesis result at time of death. The mean TESS was 73% (29–93%) at last follow-up
Discussion: Although this is a small series, the long-term results appear to indicate that reconstruction of the lower extremity of the humerus with a massive prosthesis is a satisfactory option for this rare tumour localisation.
The abstracts were prepared by Docteur Jean Barthas. Correspondence should be addressed to him at Secrétariat de la Société S.O.F.C.O.T., 56 rue Boissonade, 75014 Paris.