Abstract
Aseptic loosening is one of the most frequent cause of total knee arthroplasty (TKA) failure; it is related to a fatigue-type wear which can rapidly break up a tibial polyethylene (PE) inlay. PE debris production is the result of this wear and depend on contact stress between the components. This crucial parameter is more related to conformity than to load tranfer. That is why mobile bearing TKA seems to represent a valid solution to the PE debris production. In fact this implant offers great tibio-femoral conformity without an increased risk of loosening due to increased axial torque. Mobile bearing TKA also eneables the surgeon to self correct tibial component malrotation. On the other hand this implant could lead to new problems such as bearing dislocation or breakage and a possible new wear pattern at the PE-tibial metal tray interface.
We performed 100 consecutive cemented posterior stabilised TKA using randomly in 50 cases a mobile bearing and in 50 cases the same implant with a fixed bearing. The mean follow up was 15 months. The results in mobile bearing group are the following. Hospital for Special Surgery (HSS) score improved from 38 to 91 after the surgery. Preoperative Range of Movement (ROM) was 71 while the post op value was 107. No lucencies > 1 mm were observed. There were no statistically significant differences among the mobile bearing group and the fixed bearing one. Despite our short follow up, this implant offers the same good clinical results if compared to other devices, and allows the surgeons to correct errors in positioning the implant.
In future we have to consider that this possibility will lead to an abnormal load stresses distribution and to a higher risk of late loosening.
The abstracts were prepared by Nico Verdoschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, Universitair Medisch Centrum, Orthopaedie / CSS1, Huispost 800, Postbus 9101, 6500 HB Nijmegen, Th. Craanenlaan 7, 6525 GH Nijmegen, The Netherlands.