Abstract
Aim: To determine the accuracy of the posterior-to-anterior (PA) 15 degrees caudad view of the clavicle to assess amount of shortening of clavicular fractures. Method: The first stage of the study involved taking x-rays of an adult skeleton, centred on the clavicle. The projections included the standard anterior-to-posterior (AP) 15 degrees cephalad view, and the PA 15 degrees caudad view. Additional images were taken in the 15 degrees caudad view with a series of oblique rotational views, and oblique images in the vertical plane. Metal markers were placed on the clavicle at 10mm intervals. The clavicular length and the interval between markers were measured on the x rays.The second stage of the study involved obtaining the PA 15 degrees caudad x-ray on 50 patients with clavicular fractures. The non injured clavicle was also x-rayed. The lengths of the non injured clavicle and the lengths of the fragments of the fractured clavicle were recorded.
Results: The length of the clavicle of the skeleton in the AP standard image was 149mm. The length in the PA 15 degrees caudad image was 130mm, with a maximum of 4mm variation on the oblique views up to 30 degrees. The true length was 124mm.Forty-five fractures were diaphyseal, and five were outer third fractures. There was less than 5mm measured difference in the length of injured and non injured clavicle in 38 out of 45 patients with diaphyseal fractures (84%).
Conclusions: The PA 15 degrees caudad clavicular x-ray provided a more accurate assessment of clavicular shortening than the standard AP view, and was well tolerated by the patients.
The abstracts were prepared by Professor A. J. Thurston. Correspondence should be addressed to him at the Department of Surgery, Wellington School of Medicine, PO Box 7343, Wellington South, New Zealand