Abstract
We have made a retrospective review of 185 cemented Charnley total hip arthroplasties performed between 1970 and 1974 to determine the relationships between radiological variables and failure of the femoral and acetabular components. We measured the acetabular wear, the orientation of the cup, the thickness and consistency of acetabular and femoral cement mantles, radiolucency and femoral alignment. The mean follow-up was for 11.7 years. Femoral loosening was demonstrable radiologically in 15 hips (8.1%), ten (5.4%) of which were revised during the period of follow-up. Only when the first postoperative radiograph showed a thin cement mantle in Gruen zone 5 was there a significant association with failure of the femoral component. There were 12 loose acetabular components (6.5%), nine (4.8%) of which were revised. When the initial radiograph after operation showed radiolucency in DeLee and Charnley zone 1, the incidence of acetabular loosening was 28.21%. If such radiolucency was not present, the incidence of acetabular loosening was only 0.69%. Our findings emphasise the importance of careful cementing.