MEASUREMENT OF POLYETHYLENE WEAR IN CEMENTLESS TOTAL HIP ARTHROPLASTY



Download PDF

Abstract

We made a clinical study of polyethylene wear in 240 hips of 187 patients having primary total hip arthroplasties from 1989 to 1990, using uncemented Osteonics components, with a head size of 26 mm. We excluded cups with anteversion of over 20° and measured linear wear by a new method using a digitiser and special software of our design. Follow-up was from two to five years (mean 4.3).

The mean age at operation was 50.3 years, with more men than women (1.4:1). The mean linear wear per year was 0.15 mm; this did not increase with the longevity of the prosthesis (p = 0.54). In 59 hips showing evidence of osteolysis, the mean linear wear rate was significantly higher at 0.23 mm/year (p < 0.001). The mean linear wear rate also correlated significantly with age at the time of operation (p = 0.008), but we found no significant correlations with body-weight, gender, aetiology of the disease, thickness of polyethylene, or cup position.

Our new method of measurement is time-saving and reproducible. The results confirm the greater rate of linear wear of polyethylene in patients showing osteolysis and in those who are younger.