Abstract
We report the results of 75 consecutive primary press-fit Kinemax arthroplasties with an average follow-up of 14 months (three to 28). We reviewed 26 cemented and 49 non-cemented tibial components implanted into 72 patients (30 men and 42 women, median age 70 years). At the latest follow-up the overall evaluation (Hospital for Special Surgery knee rating scale) for cemented cases was excellent in 54%, good in 42% and poor in 4%. No cemented prosthesis loosened. Of the non-cemented cases 55% were excellent, 23% good, and 2% fair. Ten tibial implants (20%) loosened and required revision. Residual pain marred the result in 24% of the non-cemented prostheses and in 4% of the cemented group. We do not recommend the press-fit, smooth-surfaced Kinemax prosthesis for non-cemented use.