Abstract
The anterior cruciate ligament was replaced in rabbits, using implants of carbon or polyester filaments with known mechanical properties. The biocompatibility of the implants was assessed in detail using light microscopy, and scanning and transmission electron microscopy. Mechanical tests were made of stability, in comparison with normal joints and controls after excision of the ligament. Some carbon fibre implants broke down in vivo, allowing instability; the fragments caused chronic inflammation. Intact carbon implants did not induce the formation of neoligaments; they were covered by tissue, but there was no ingrowth. Polyester did not degrade mechanically and supported early collagenous ingrowth within the implant, even in the mid-joint space. It was concluded that there was no justification for the use of carbon fibres as anterior cruciate replacements; polyester appeared to be suitable.