Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Socket design and cement pressurisation in the Charnley low-friction arthroplasty



Download PDF

Abstract

An experimental evaluation was made of the efficacy of an unflanged cup, an ogee-flanged cup and the Exeter pressuriser in the pressurisation of acetabular cement. Only a very modest injection pressure could be generated through an unflanged socket and even this was readily lost as the socket "bottomed out". The ogee-flanged socket gave a consistently high injection pressure which could be maintained throughout the process of polymerisation. The Exeter pressuriser gave marginally better results. Of the two cements tested, Palacos with gentamicin showed better intrusion than CMW Type 1 cement, but it also allowed greater extrusion once pressure was released. The importance of maintaining a continuous pressure on the cement throughout polymerisation is emphasised.

For access options please click here